Sterilisation Flashcards
What is the difference between a terminally sterilised product and an aseptically processed product?
Terminally sterilised is where the final product is heat treated in its final container on a validated sterilisation cycle. Aseptically processed product cannot be heat treated without spoilage of the product, it will undergo a sterilisation step such as sterile filtration as part of the manufacturing process then then any subsequent steps need to be strictly controlled as to not introduce any contamination of the product
Can you tell me about the different methods that can be used for sterilisation?
Moist or dry heat, depyrogenation, gamma radiation, Ethylene oxide, VHP, sterile filtration
How would you monitor autoclave performance on a batch by batch basis?
I would review the sterilisation trace and look at critical values such as temperature, pressure and time. A master control record can also be used to compare the batch trace is as per expected
What other routine tests or checks would you perform on an autoclave?
Leak rate and Bowie dick
What does a Bowie-Dick test tell you.
Steam penetration, it should produce an even colour change to the indicator paper located within the pack
Validation of an autoclave. Describe.
URS, Commissioning and cycle development, IQ - document machine info including software and firmware versions, drawing walkdown, component installation, calibration, etc, OQ - Leak rate, bowie dick, air detector test, automatic control tests, thermometric small load, PQ - include min and max loads for each required load and includes BIs
Tell me how you would test for steam quality and is there a guide?
ISPE baseline guide and BS EN285 should be used for the validation of autoclaves. Steam quality testing includes dryness fraction, superheat and non condensable gases
What would you do if, after a routine re validation, a BI was found to be viable?
I would follow the OOS procedure to confirm it was a valid result. If confirmed as valid I would raise a deviation and work through - scope and contain - stop production and release and block anything in my control then quickly look into the root cause as this is a routine re validation I would want assurance that anything released since previous successful validation was not impacted. I would look at the cycle from the failed BI run and see if there was anything unusual from the prep of the load, dry load on completion, load pattern as defined etc. Has there been any maintenance carried out, if so what and when etc and use this information to reduce scope
What do you understand by the term SAL
Sterility assurance level is the probability that a single unit that has been subjected to sterilization nevertheless remains nonsterile. It is never possible to prove that all organisms have been destroyed, as the likelihood of survival of an individual microorganism is never zero.
Can you differentiate between the environmental conditions of terminal sterilisation and aseptic sterilisation.
Open manipulations and filling etc in aseptic manufacture should be carried out in a grade A zone where as terminally sterilised product such as a sterile nasal spray for example can be carried in is less stringent conditions such as Grade C
What do you understand by Fo? What are the units? What is the minimum Fo?
Fo value is the time equivalent to 1 minute at 121oC. F8 is the minium Fo value expected = needs at least 8 minutes at 121oC
What would be the effect if the steam was super saturated?
Superheated steam is too dry so doesn’t have the same kill as moist steam - takes a lot longer to achieve the same result
What is the D value?
Time taken to see a 1 log reduction
What is the Z value?
Temperature increase required to effect 1 log reduction
Is there anything you should consider during autoclave validation where the autoclave will be used for both porous and non-porous loads?
Cycle development and validated loads
Autoclave validation. Air detector, what design parameters?
An air detector is fitted to certain sterilizers that employ vacuum as a means of removing air from the load before sterilization. It is required for porous load sterilizers. The air detector should cause a fault to be indicated if the amount of air or gas in the chamber at the start of the plateau period is sufficient to depress the temperature in the centre of the load more than 2ºC below the temperature in the active chamber discharge.