Spirometry calibration Flashcards
define quality control
a series of checks to ensure that results generated are as accurate and precise as possible
-need to be good repeatability (during the test) and reproducability (between visits)
What is calibration?
-comparing the measurements of two instruments, one with a known magnitude or correctness (standard device) against which you measure the other and correct if necessary
What is verification?
checking an instrument and ensuring any error of measurement is within permissible limits
-spirometers have their calibration verified by injecting a known volume, 3L syringe is used
What value is acceptable for verification?
+/- 3% of true volume
Why do we calibrate a spirometer?
-Spirometers are precision instruments which have been calibrated by the manufacturer.
-If the accuracy is not checked, then errors would go undetected.
-Correct interpretation of spirometry can only be made if the results are accurate.
-By performing spirometry to ARTP standards, you can have confidence in the results produced.
Describe the verification of calibration process
Use a 3 litre syringe.
-it must read +/- 3% of true value i.e. between 2.91L – 3.09L.
-Calibration should be verified every morning before each clinic session and again after every 4 hours of use.
-Recheck if room temp changes by more than 5oC
-A calibration log should be maintained (software or logbook)
-Calibration syringe need to be checked YEARLY
What should you do if the verification of calibration is unsuccesful?
-return the spirometer to the manufacturer for recalibration
-can’t be used for testing until it can be successfully calibrated
Why do some spirometers need calibration checks at different flows?
-low, middle and high flow rate are generated by applying different pressure to the calibration syringe
-checking each flow is getting the volume
How do you chek the physiological control of a spirometer?
-perform 10 consecutive spirometry tests which should be reproducible and technically acceptable .
- performed by the same operator at a similar time of day.
-Add each of 10 FEV1, 10 RVC, 10 FVC, and 10 PEF together and divide by 10 to get average.
-Work out 5% either side to represent normal variability and create acceptable operating range.
-Check your own against the range at least weekly and log results.
Not to be done instead of the calibration syringe but as an extra level of calibration
What needs to be included in the calibration log book?
-equiptment calibrated(model, serial number)
-date
-time
-temp
-operator
-syringe used
-was calibration check successful?