SPECT/CT and registration Flashcards
1
Q
Benefits of CT
A
- Hybrid systems combine functional &
anatomical imaging in a single study - CT significantly improves
- Attenuation correction
- Anatomic localization
2
Q
Attenuation correction CT
A
- Pixel values require transformation
- Hounsfield units (HU) to attenuation coefficients
- CT number = [(μt – μw) x k] /μw
- μt = LAC of tissue
- μw = LAC of water
- k = 1000 (constant)
- Adjusted for energy differences between photons being imaged and CT
3
Q
Image Co-Registration Steps
A
- Decide on the role for the data sets
- Decide on the type of information used to register
- Define the allowed transformation
between the model and the reference - Define the goodness of match criteria
- Perform optimization
4
Q
Forms of registration
A
- Interactive
- Landmark Based
- Surface Matching
- Automatic
- Maximization of Mutual Information
- Elastic
5
Q
Instrumentation and QA
A
- Gamma Camera
- CT Scanner
- Operation of each scanner as an individual piece of equipment is possible
- QC and preventative maintenance (PM) as it applies to each scanner is required
6
Q
Clinical Applications
A
- Myocardial Perfusion Imaging
- 99mTc MDP
- 111In WBC
- 99mTc HMPAO WBC
- 131I WBS
- 123I MIBG
- 111In Octreotide
- 67Ga