Attenuation Correction Flashcards
1
Q
Photon attenuation
A
- absorption and scattering interactions
- degree of attenuation depends on: photon energy, thickness of tissue, linear attenuation coefficient
2
Q
Attenuation coefficients
A
Linear attenuation coefficient
* Symbolized by μ or μl
* Expressed in cm-1
Mass attenuation coefficient
* Symbolized by μ/p or μm = μl/p
* Expressed in cm2/g
3
Q
Narrow-beam geometry
A
- Narrows acceptance of scattered photons
- Requires narrow aperture
- Absorbed halfway between source and detector
4
Q
Broad-beam geometry
A
- Allows acceptance of scattered photons
5
Q
Photon depth
A
- attenuation increases with depth
- apparent decrease in deep vs. superficial activity
6
Q
Attenuation correction
A
- AC can be applied before, during, or after reconstruction
- methods: arthmetic mean, geometric mean, chang, transmission (ct, rod source)
7
Q
Arithmetic and geometric mean
A
- correction based on the path length of photons from origin to the body surface
- opposing views are combined using arithmetic mean or geometric mean
- 360 acquisition
- a constant attenuating medium is assumed
- AC factor is applied prior to reconstrution
8
Q
Chang method
A
- body contour defined by ROI
- AC factors calculated based on distance from ROI, then applied to each pixel
- a constant attenuating medium is assessed
- performed post-reconstruction
9
Q
Transmission methods
A
- an external source of radiation is used to measure tissue attenuation
- radionuclide and CT options
- transmission scan generates an AC map
- AC factors based on the LAC of the tissue
- AC map is applied to the emission data to correct for attenuation
- applied prior to FBP and incorperated into IR
10
Q
RN Tranmission Scans
A
- Blank/reference scan performed prior to transmission scan
- Transmission and emission scans are acquired using the same detector system
- Simultaneous vs. sequential scans
- Transmission scan performed with flood source, line source(s), or moving line source
- Typically, long-lived sources (e.g., 153Gd)
- Must take the difference between photon energies into account
11
Q
CT transmission scans
A
- Best method of AC in current practice
- Transmission and emission scans are
acquired sequentially - CT coefficients need to be adjusted to
reflect the difference in x-ray and photon
energies
12
Q
Attenuation artifacts
A
- Without AC, artifacts include:
- Hot rim
- Areas of decreased uptake
- False positives/perfusion defects may result
- With AC, artifacts include:
- Motion
- Truncation
- Contrast/metal
13
Q
Scatter correction
A
- Scattered photons of up to 50 degrees have up to a 50% chance of acceptance
- Scatter will increase the counts in the ray sum
14
Q
Method 1 scatter correction
A
- use asymmetric window
15
Q
Method 2
A
- Compton scattering window subtraction
- Uses two PHA windows
- Subtract the second window from the first, then reconstruct