Selectra Flashcards
Selectra outer catheter french sizes
9 outer
7 inner
Selectra inner catheter inner french
5F inner with 7F outer
Vertical CS takeoff is associated with what
Enlarged heart
How many outer catheter shapes are there
8
How many inner catheter shapes are there
2
Minimum acceptable QLV
80-90ms
What is the break point at which remodelling stops
6 months
Factors associated with less remodelling
Male, Ischemia, QRS <140ms
Updated outer and inner catheters now have what
Reinforced segments near hub 20mm
Updated outer catheters now have what
Luer-Lock connectors
Tuohy-borst connector
What does the new TVI tool accomodate
Things up to 7F - lead + wire inserted
less likely to be sucked into catheter
What does the Tuohy-borst adapter offer
No bleed back / not required for implant but an option
Lead can’t be passed through
Only needed with .014 wire
What two lengths can I get selectra outer catheters in
45cm / 55cm
What two lengths can I get selectra inner catheters in
59cm and 69cm