Science Of Energy Flashcards
Advanced bipolar instruments incorporate the following improvements over traditional bipolar instruments:
Impedance sensing, increased compression, mechanical cutting, 7mm vessel seal
“Energy always seeks the path of least resistance.” This principle relates to which important tissue effect?
Lateral thermal spread
What are the controllable factors that impact tissue sealing? (Select all that apply)
Heat, time, compression
What surgeons sometimes do to coagulate vessels> 1mm with a monopolar pencil?
Use a metal instrument (forceps or hemostat) to apply compression while using intentional direct coupling
Which of the following is NOT benefit of harmonic technology?
Faster than monopolar
Vessel Sealing up to and including 7mm is a benefit of increasing the physical factor in advanced bipolar devices.
Compression
Ohm’s law states that voltage is equal to current times _____.
Resistance
As tissue dedicates (dries out), the ____ within the tissue ______, causing energy to move to a new path outside of the intended area.
Resistance, increases
What is NOT true about monopolar devices?
Monopolar devices require longer time on tissue, compared to bipolar or ultrasonic devices
When using monopolar pencils, which action below may decrease the risk of a surgeon getting burned when intentionally using direct coupling (“buzzing” a metal instrument to add compression”)
Holding the instrument in a way that provides more surface area and contact, to minimize concentration
Nerve and muscle stimulation ceases at ____ kHz, because the _____ is so rapid across the body’s neurons that the brain doesn’t recognize it.
100 kHz, frequency
Which statement below is FALSE?
Monopolar devices use direct current and bipolar devices use alternating current
Which of the following energy modalities requires a return pad?
Monopolar
The smaller the area of contact, the greater the current concentration, resulting in greater tissue effect. TRUE OR FALSE
TRUE
What is the relationship between current concentration and time on the tissue?
Higher current concentration and longer time on tissue means more thermal tissue damage