Safety Flashcards
What needed to be checked after connecting to your patient?
- Assess patient and ventilator system
- -> synchrony
- -> alarm is appropriate
- ->waveform
- ->patient vital signs
2, ensure tidiness and safety of circuit
- BVM, intubation tray close by
What are the type of high and low pressure alarm?
Circuit pressure (set 5 - 10 cmH2O above PIP)
Ventilator pressure
Input pressure
Which PB840 alarm is always active?
high circuit pressure alarm
When does the severe occlusion alarm go off in PB840
- The patient circuit is severely
occluded
- The ventilator enters occlusion status cycling
- ->by opening the safety and exhalation valve (so that patient can still breath)
The
elapsed time without
ventilatory support is displayed.
What happen to PB840 when the high pressure alarm goes off
it will cycle to peep
What is the result of ventilator inoperative?
- cannot use vent until maintenance is done
- safety valve will open
- could be a hardware or software problem
- need to replace ventilator
When does the safety valve open?
- Both gas supplies lost
- Severe Occlusion
- Vent Inop
- System pressure > 100 cmH2O
What is the purpose of the Disconnect sensitivity of PB840?
it detects a disconnect when the expiratory pressure transducer measures no circuit pressure and no exhaled flow during the first 200 ms of exhalation
- used for VT discrepancies (such as disconnect between wye and patient
What does the Dsens monitor
Expiratory pressure
Expiratory flow
Delivered volume
Exhaled volume
What is the function of the apnea interval (TA)?
- an alarm that goes off when there is no breath trigger by patient or ventilator following the TA interval
What does the TA reset?
Alarm will reset if two patient breaths are detected, with > ½ of original VT
How is apnea interval measured?
TA is measured from the start of an inspiration to the start of the next inspiration and is based on the ventilator’s
inspiratory detection criteria
what happens if the Paw high alarm of evita goes off?
upper alarm limit for the airway pressure has been exceeded. The
device immediately reduces the airway pressure to the set PEEP. If this
is not successful and the upper alarm limit is exceeded by 5 cmH2O,
the airway pressure is reduced to ambient pressure.
What is the hose kinked alarm about in Evita?
inspiratory pressure exceeding 30 cmH2O is measured for more
than 30 seconds during O2 therapy.
How does peep high alarm go off in EVita?
- The measured PEEP is higher than the set value:
– higher by 8 cmH2O for 2 ventilation cycles or 15 seconds
– higher by 5 cmH2O for 10 ventilation cycles