RKI eagle Flashcards
What is alarm 1 factory setting?
Lows alarm (10% LEL, 19.5O2, 10H2S, 25ppmCO
Alarm 2 factory setting is a ___ alarm
High alarm (50%LEL, 23.5 O2, 30ppm, 50ppm
Sampling method is ?
Sample draw
Response time is
T90 within 30 seconds
What type of sensor? For gas
Catalytic
The eagle 2 uses up to ___ sensors
6 to monitor O2 LEL H2S CO
Alarm LEDs ___ sets of red alarm LEDs one in the left and right corner will alert you to gas, low battery and failure alarms.
Two
How many and what size batteries
4c alkaline batteries
A ___ ft polyurethane sample hose and ___ inch hydrophobic probe are included as standard
5, 10
A Teflon hose is provided with all units
PID sensor
What will you see for a failed sensor on the LCD display
XXX
Fresh air in a clean environment. Press and how the ____ ___ _____ and the buzzer will pulse while you hold the button
Air ^ yes
Probe only sample time
12sec
Probe only sample time
12sec
Probe plus 50 ft Jose =
35sec
Probe plus 75ft
45 sec
Probe plus 125ft
75sec
The eagle to is calibrated to what
CH4 or methane
250ppm increments from
10,000 ppm to 50,000ppm
False alarms may be caused by what two things
Radio frequency or electromagnetic interference EMI
What will appear on the screen and audibly when meter is over range
OVER appears next to the gas reading LED flashes twice per second and pulsating tone twice per second
Low battery will present with
BATT along the left side of the LCD
Sensor failure will appear as
Failed sensors at the top of the display
What causes flow failure
Probe tube is clogged/hose has a kink
Display indicates SYSTEM FAIL 12 during start up. What’s the cause?
Memory error has occurred
Operating range is
125ft