Minirae3000 Flashcards
Uses PID with. 9.8 ev 10.6 ev Or 11.7 ev ____ discharge lamp
Gas-discharge lamp
____ hours of continuous monitoring with rechargeable battery pack
16
____ point datalogging storage capacity
260,000
4.2 V/ 3300mAH ____ battery ____ batteries
Lithium ion battery, 4AA BATTERIES
Less than ____ hrs full charge
8
—- hours continuous
16
Operating in ___ mode
Hygiene and search
Humidity is ___ to ___
0% to 95%
Measurement range for 10.6 ev is
0.1ppm to 15,000ppm
Response time is
2 seconds
Flexible ___ “ tubing
5
___ VOC gasses
200
Accuracy for isobutylene __ to ____ ppm
10 to 2000
Sampling pump flow rate is ___ to ____
450 to 550
Flashing red LEDs indicate
Exceeded preset limits, low battery, or sensor failure
Turning off the instrument hold the ___ key for 3 seconds a 5 second countdown to shutoff begins
Mode
Operation mode
Hygiene
Pump that provides a 450 to 550cc per minute flow rate l. This pump can pull in air samples from ___ away horizontally
200’
This meter is calibrated to
Isobutylene
Message: HIGH, OVR, MAX, pump icon flashes, lamp, ___ beeps 1 flashes per second
3
Message: TWA, STEL, CAL, NEG, battery icon flashes _ beeps 1 flash per second
1
What is the max time for one event
24hrs 28, 800 points
How many hours of operation will alkaline batteries have
12
What is the external filter made of?
Teflon
How do you enter calibration?
Press and hold mode and N/-
Under basic user level you
Do not need a password
Zero calibration starts a ___ second countdown and displays what message
30s, and zeroing
When span calibration is done you’ll see thus message and ppm reading
Span 1 is done! 100.0ppm
What is the default password?
0000
STEL is expressed in
PPB
How to change the gas measurement or correction factor?
Scroll to ppm second one in
How to get to the datalog
Fourth to the right (paper looking symbol)
The default value is 60s what is the max?
3600
What three types of data can you choose?
Average, max, min
The time is in
Military
When powered on the instrument
Zeros itself
In hygiene mode measurements are continuously running in search mode
It is manual start and stop
LCD contrast min and max value
20/60
How many screws are on the battery compartment
3 Phillips head
What are major maintenance items of the instrument
Battery pack, sensor module, PID lamp, sampling pump, inlet connectors and filters
When cleaning the PID sensor it is highly recommended that ____ for 15 minutes
An ultrasound bath is to be used to clean the sensor
What causes an abnormally high or low reading
Dirty filter or sensor
What key do you press and hold to turn on?
Mode
Low voltage alarm you should
Turn off the instrument within 10 minutes