RF Genorator (Smartablate) Flashcards
Catheter port 1 connects
Smartablate generator to PIU RF Generator Port
Relays ablation data via smartablate generator cable (red/grey, white/grey)
Note: if non NAV RF cath, cath plugs directly into this port
Grounding pad/indifferent electrode port
Start with port 1
Completes RF circuit connecting patient to the RF Generator
ECG port
Connects RF to a recording system when using our Celsius catheters
Irrigation pump port
Back of generator
Connects generator to smartablate irrigation pump
Blue port, blue redel cables
IOIO 1
Connects RF generator to EP Recording system
IOIO2
Connects RF generator to workstation relaying ablation data to Carto 3 system
Back of workstation
Foot pedal port
Connects foot pedal to RF generator. Control ablation with pedal
Remote Control Port
Connects RF generator to RF Generator remote panel
Yellow port, yellow redel cable
Start new procedure
New patient icon (bottom of screen)
Yes
Values reset
Ablation screen
Displayed by default
Time, flow, target, temp, and target power can be adjust on the screen at anytime
Catheter selection
Bottom of screen
Irrigation screen
Bottom panel right side
Input irrigation bag size
Reset bag when changed
Flash flush from generator/remote
2 modes the user can choose
Power control
Watt control
Power control
Physician controls power manually.
Dictate a specific amount of watts to use
Temperature does not impact the amount of watts delivered
Temperature control
Physician chooses a desired temperature. Generator delivers enough watts to maintain that temp
Temp and watts change continuously as generator attempts to maintain set temp