What gases use INFRARED analysis on the multigas analyzer?
- CO2
- N20
- Anesthetic agents
What gases use PARAMAGNETIC analysis on the multigas analyzer?
O2
Why do we monitor respiratory gases?
Capnography
Breathing circuit problems
End tidal AA – estimate alveolar concentration
Adequacy of pre-oxygenation
Anesthesia machine failures/malfunctions
Physiologic events - bronchospasm, etc.
Gas supply problems
Gas uptake and production
What is the SIDE STREAM gas sampling technique?
Gas withdrawn from the airway and diverted to the analyzer.
What are the characteristics of the SIDE STREAM technique? (2)
diverting analyzer
withdrawal sampling
What is the rate and the response time for the side stream technique?
50 - 250 ml/min
300 ms
What is the MAIN STREAM gas sampling technique?
Infrared light beam is passed through the airway.
There is more dead space when the sample line is (before/after) the HME.
after
What does the circle depict?
The sensor has been separated from the cuvette, which contains the window through which infrared light passes.
What changes in monitoring will occur when the sample line is behind the HME?
decreased Et CO2
What is a disadvantage of attaching the sample line on the patient side of the HME?
water and bacterial/viral contamination of the sample line and then the monitor
What can occur if the water trap is located under the vaporizer?
Plastic that makes the trap can be damaged if contacted by a liquid anesthetic agent
What can occur if the gas sampling line is attached to the breathing circuit near the Y-piece?
Fresh gas entering the inspiratory limb will dilute readings, thus place the gas sample nearest to ETT
What is the volume of gas that is sampled?
2 - 5 ml
It is not useful to measure EtCO2 with loose fitting masks or nasal cannulae. T or F?
False
The presence of exhaled CO2 provides info that the pt is breathing even though the values will be diluted with oxygen and air.
There are automatic and manual agent detectors.
If an agent is erroneously detected, what should you do to correct the error?
Turn the machine off then back on.
What is the benefit of an automatic agent identifier over a manual one other than the obvious answer?
Automatics can detect contaminated vaporizers.
What are the techniques used to analyze oxygen?
What is the fastest?
Polarographic
Fuel cell
Paramagnetic (is the fastest that we need to know)
Mass spectrometry–technically the fastest
What is the lifespan of a fuel cell? (2)
What is the response time?
200,000% hours –> 2000 hrs
~15 months in RA
5-35 sec
The paramagnetic technique has what response time?
150 msec