quality control Flashcards
1
Q
centering
A
center of xray beam must be aligned with indicator within 1% of SID
2
Q
distance
A
actual SID must be within 2% of SID
3
Q
collimation
A
light field must align with xray field collimation within 2% of SID
4
Q
exposure timer
A
actual exposure time must be within 5% of indicated time
5
Q
AEC testing
A
multiple exposures through different thicknesses should produce the same exposure amount
6
Q
kVp
A
actual kVp must be within 5% of assigned kVp
7
Q
exposure reproducibility
A
exposure values for identical technical parameters should be within 5%
8
Q
exposure linearity
A
exposure per mAs should not change more than 10% between mA stations