QM QA QC Flashcards

1
Q

the attributes of a product or service offering its specific attributes to a degree of excellence

A

definition of quality

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

evaluation of different processes or tasks that occur in order to perform your job well

A

quality improvement

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

things that managers need to assure are present in the department, such as qualified technologists

A

quality assurance

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

the testing of equipment and accessories to ensure that they are working well

A

quality control

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

what is a quality radiograph?

A

one that posesses sufficient density, adequate penetration, appropriate contrast, defined structural outlines and resolution between tissue opacities

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

what must be developed to assess if a product or service meets, fails, exceeds standards?

A

standard criteria

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

an image density of ___ on anatomy of interest is a standard to evaluate density

A

1.0 DU

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

the extent to which the service provided is expected to meet the standard criteria and achieve the most favorable result

A

degree of quality

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

accepted standard deviation of density

A

20% or .2 DU above or below

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

goal of quality management

A

zero defect

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

zero defect

A

free of any flaws or deficiencies

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

where do non-technical characteristics prevail

A

determination of quality to customers, communication methods, degree and standard of patient care, cleanliness and design of facility’s environment

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

in order for a facility to measure quality of services, it must define and understand

A

its purpose, responsibilities, and objectives

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

how does a facility define and understand purpose, responsibilities, and objectives?

A

defining vision, mission, business values, and guiding principles

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

mission statements are designed to provide

A

the organization, employees, and customers with a true understanding of the operation and ultimate goal of the organization

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

the understanding of what resources the business must develop and value to success

A

business values

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

provide the values and strength for an organization to work during times of change

A

guiding principles

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

A management responsibility to assess safe and proper performance

A

Quality assurance

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

In addition to quality control, quality assurance includes:

A

preventative maintenance, acceptance testing, education and training

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

The development of equipment specifications (list of features) is part of

A

quality assurance

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

Which of the following persons must be involved in a Quality Management program at an imaging facility?

A

administrative director, medical director, technologists

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

Quality assurance programs are important, because they enable departments to

A

maintain a minimum of education / training qualifications, reduce unnecessary wastes and associated costs, improve technologists’ performance

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

Quality Improvement programs include quality assurance, quality control and:

A

evaluation and improvement of all clinical processes

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

Quality can be defined as

A

attributes or characteristics of a material at a specified level

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
Q

Zero defect refers to which of the following conditions?

A

free of any deficiencies

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
26
Q

Which of the following expectations are often expected and not evaluated by customers?

A

technical skill

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
27
Q

A mission statement for an organization should identify which of the following?

A

all

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
28
Q

One purpose of a quality management team is to

A

encourage and necessitate formal discussions of quality issues

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
29
Q

Which of the following are not a part of the quality improvement evaluation process?

A

Direct

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
30
Q

When evaluating the services a facility provides, the quality management plan should include

A

service attributes, the technical work, interpersonal skills

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
31
Q

A series of regular detailed inspections, checks and calibrations of equipment’s electrical and mechanical components to ensure the equipment is operating properly, accurately and safely is referred to as:

A

preventative maintenance

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
32
Q

A list of desired features of a new imaging unit used to solicit purchase bids is referred to as:

A

equipment specifications

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
33
Q

The evaluation of the effectiveness of report generation is part of which of the following?

A

quality improvement

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
34
Q

After a new unit is installed and before use, which of the following should be performed?

A

acceptance testing

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
35
Q

The management of staffing qualifications and training are generally part of which of the following programs?

A

quality assurance

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
36
Q

Which of the following would not be considered under quality assurance?

A

work flow productivity measurements

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
37
Q

Which of the following is not a benefit of a well organized and implemented quality management program?

A

greater staffing levels

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
38
Q

Quality Control can be best defined as

A

A set of tests or tools that measure equipment performance

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
39
Q

Testing to check mA linearity is a component of which of the following activities?

A

quality control

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
40
Q

Which of the following areas of the radiographic imaging chain should be routinely checked to ensure safe and proper operation?

A

cassettes and grids, radiographic processor, the darkroom environment

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
41
Q

Dust and dirt in a darkroom can be more easily visualized under which of the following conditions?

A

blacklight

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
42
Q

When assessing the operation and integrity of view boxes, an R.T. should evaluate?

A

consistency in color of bulbs, the color and consistency of white plexiglass covers, consistency in illumination between boxes

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
43
Q

When measuring the daily sensitivity of film processing which of the following are evaluated?

A

base plus fog, Speed index (MD), Contrast index (HD)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
44
Q

To test the congruency of the collimator light or collimator alignment, an R.T. would perform which of the following tests?

A

nine penny test

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
45
Q

Congruency between the light field and the image receptor should not be more than what percent off?

A

3% of the SID

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
46
Q

Reciprocity testing measures:

A

consistency of MAS values

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
47
Q

MA linearity testing measures:

A

the doubling effect of mA stations

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
48
Q

When testing viewboxes for light intensity, a photometer measures the light emanating from the surface of the viewbox cover. What is the preferred light intensity for radiography?

A

500 candelas (ft-c)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
49
Q

The spinning top test can be used to measure:

A

timer accuracy

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
50
Q

If an R.T. does not have the access to a sensitometer, the R.T. could use, which of the following devices to make a reproducible sensitometric image everyday?

A

penetrometer

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
51
Q

The CONTRAST index is the difference between the step closest to 2.2 but not over and the step closest to _______ but not under.

A

.45

52
Q

Wisconsin Test Cassette or a Argon Crooke’s tube can be used to measure which of the following aspects of x-ray production?

A

kV accuracy

53
Q

A term used to describe the alignment between two devices, such as grid and tabletop is:

A

congruency

54
Q

When inspecting the integrity of a processing unit, the R.T. must check:

A

integrity of electrical wiring, the gears and solution tank condition, spillage and cleanliness

55
Q

Testing to check kV accuracy must not deviate more than:

A

5%

56
Q

Repeat Image analysis provides information about:

A

understanding of exposure factor standards, an individual’s technical performance, equipment problems

57
Q

A MANAGEMENT responsibility to assess safe and proper performance.

A

Quality Assurance

58
Q

Quality Assurance includes

A

PM
Equipment Specs
Acceptance Testing
Education and Training
Quality Control

59
Q

Development of equipment specs is part of what?

A

Quality Assurance

60
Q

Evaluation and improvement of all clinical processes

A

Quality Improvement

61
Q

attributes or characteristics of a material at a specified level

A

Quality

62
Q

Intent/importance of of organizational goal and how it will be reached
degree of commitment
products and services will provide
market products will be delivered
scope of each individual

A

Mission Statement

63
Q

Quality Management Program should consider

A

the technical work
medical care provided
interpersonal relationships
the physical environment
service attributes.

64
Q

A series of regular detailed inspections, checks and calibrations of equipment’s electrical and mechanical components to ensure the equipment is operating properly, accurately and safely is referred to as

A

Preventative Maintenance

65
Q

A list of desired features of a new imaging unit used to solicit purchase bids is referred to as:

A

Equipment Specifications

66
Q

The evaluation of the effectiveness of report generation is part of which

A

Quality Improvement

67
Q

After a new unit is installed and before use, which of the following should be performed

A

acceptance testing

68
Q

The management of staffing qualifications and training are generally part of what program?

A

quality assurance

69
Q

A set of tests or tools that measure equipment performance

A

quality control

70
Q

Testing to check mA linearity is a component of what

A

quality control

71
Q

To test the congruency of the collimator light or collimator alignment, an R.T. would perform which test

A

nine penny

72
Q

Reciprocity testing measures what

A

mAs value consistency

73
Q

mA linearity measures what

A

the doubling effect of mA stations

74
Q

The spinning top test measures what

A

timer accuracy

75
Q

Wisconsin Test Cassette or a Argon Crooke’s tube can be used to measure which aspects of x-ray production?

A

kV accuracy

76
Q

Definition of quality must include

A

-be all- inclusive
-describe every feature or attribute,
-describe any other criteria that it must possess so as not to hinder its ability to possess the level of quality sought and perform the job it was intended to do.

77
Q

This must:
-be stimulating and achievable
-it should define quality as it relates specifically to the facility, its ultimate goal
-instill the importance of this goal and a sense of belonging to the organization, commitment of the organization as well as the expectations of customers and vendors

A

Vision

78
Q

Mission Statement must include

A

-intent of organization
-degree of commitment of org
-products & services offered
-market to which products will be delivered
-scope of each responsible individual

79
Q

What resources the business must develop and value to succeed

A

Business Values

80
Q

Provide values and strength for an organization to work through during times of change- provides security and empowerment

A

Guiding Principles

81
Q

This includes: In-service Education, Preventative Maintenance, Equipment Specifications, Acceptance Testing, Quality Control Testing, & Quality Improvement Initiatives

A

Quality Assurance

82
Q

Thorough inspection and performance testing

A

Type 1 PM

83
Q

Equipment condition and integrity survey

A

Type 2 PM

84
Q

Equipment operation and safety inspection

A

Type 3 PM

85
Q

Stepwedge is also referred to as a

A

Penetrometer

86
Q

Processor Water tested how often

A

quarterly

87
Q

Processor time tested how often

A

monthly

88
Q

darkroom lights tested how often

A

biannually

89
Q

darkroom environment tested how often

A

monthly

90
Q

Image quality repeat assessment tested

A

monthly

91
Q

All imaging equiptment & Radiation protection devices checked

A

annually

92
Q

All accessory equiptment checked

A

biannually

93
Q

hypo retention of 2 ______ _______ is detrimental to the life of the image

A

clearing agents

94
Q

hyporetention is evaluated how often

A

biweekly

95
Q

Standard deviation of densitometer

A

+/- 1%

96
Q

Standard deviation of film design/ stock storage conditions

A

+/- 5%

97
Q

Standard deviation of sensitometer

A

+/-10%

98
Q

Standard deviation of processing conditions

A

+/- 5-10%

99
Q

Standard deviation of film characteristic indexes/ measurements

A

+/- 15% for Radiography
+/-10% for Mammo

100
Q

increased speed index = ______ Density

A

increased

101
Q

decreased speed index = _______ density

A

decreased

102
Q

increased contrast index= ______ contrast and _____ density

A

higher, higher

103
Q

decreased contrast index = ______ contrast and ______ density

A

lower, higher

104
Q

increased gross fog= ______ density

A

increased

105
Q

decreased gross fog= ______ density

A

decreased

106
Q

standard deviation of radiation output, mA timer , reciprocity and linearity

A

+/- 10%

107
Q

standard deviation of kV output & timer accuracy

A

+/-5%

108
Q

mAs reciprocity tests what?

A

the relationship between mA and time at similar mAs values when kV is constant

109
Q

mAs linearity measures what?

A

Linearity is the measure of a generator’s ability to produce sequential increases in radiation output as mAs changes incrementally. As mAs doubles, radiation output should double.

110
Q

Focal spot blooming can cause what image quality?

A

unsharpness

111
Q

What test is done to evaluation focal spot blooming?

A

Pinhole camera or star pattern test

112
Q

lack of hydroquinone will cause what image quality (reducing agent)

A

lack of blackness on image

113
Q

a depletion of sodium carbonate will produce what type of image (activator)

A

underdeveloped/ light image

114
Q

tool used to measure specific gravity

A

hygrometer

115
Q

mammo view boxes are checked how often

A

weekly

116
Q

what does the wire mesh tool test?

A

film screen contact- any spacing or air gaps will cause flux in density

117
Q

Nine penny test can be used to test what 3 things

A

light field congruency, beam alignment perpendicularity and X-ray beam bucky tray alignment

118
Q

Back up time/ max exposure should not exceed

A

500 mAs or 5 seconds

119
Q

Patient thickness and chamber selection testing density range

A

+/-0.2 DU

120
Q

Field size changes (collimation) testing density range

A

+/- 0.1 DU

121
Q

Spinning top test measures what?

A

timer accuracy

122
Q

standard deviation for timer accuracy

A

+/- 5% for times greater than .1sec and +/- 20% for times less than .1sec

123
Q

Light field congruency standard deviation

A

within 2% of SID

124
Q

Beam alignment perpendicularity standard deviation

A

within 1 degree of perpendicular

125
Q

xray beam bucky tray alignment (central ray congruency) standard deviation

A

within 1% of SID

126
Q

SID indicators measured with tape measurer standard deviation

A

+/-2% of SID