QC Flashcards

1
Q

Centering- SID

A

Actual center of the x-ray beam must align to the centering indicator by 1% of the SID

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Distance- SID

A

Actual SID must be within +/- 2% of the indicated SID

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Potential effects of the SID not being what the indicated SID is:

A

-Incorrect technique
-Increased patient dose
-Overexposed on IR

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Collimation- SID

A

The light field must align with the x-ray field collimation +/- 2% of the SID

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Exposure Time and AEC

A

Actual exposure time must be within +/- 5% of indicated exposure time

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

AEC Testing

A

-Done by making exposure using AEC on different thickness of aluminum
-Using the same technique and AEC by adding aluminum layers you should get the same receptor exposure

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

kVp Calibration

A

The actual kVp must be within +/-5% of assigned kVp

Actual kVp>assigned kVp= Increased patient dose=decreased image quality

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

kVp Reproducibility

A

(Exposure 2-Exposure1)/Exposure 1=Percentage

-Exposure values for identical technical parameters should be within +/-5%
-Same mA, Same kVp, Same exposure time

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Exposure Linearity

A

Microgray/mAs
then
(Exposure 1 mAs-Exposure 2 mAs)/Exposure 1 mAs= Exposure linearity

-Exposure per mAs(microgray/mAs) should not change more than 10% between mA stations
-When we change one mA to the next on the machine results should not be different by more than 10%