Primary FRCA Course Measurement and Monitoring Exam Prep Questions Flashcards
In pulse oximetry:
The SpO2 is determined by the absorbance of light of wavelengths 660nm and 940nm
True. Pulse oximetry relies on measuring the relative absorbance at two wavelengths (660nm and 940nm). Reduced haemoglobin has greater absorbance at 660nm and oxyhaemoglobin greater at 940nm. The ratio is the determined and the SpO2 is then calculated
In pulse oximetry:
Methaemoglobinaemia causes the SpO2 to approach 85%
True. Carboxyhaemoglobin causes an increase in SpO2 equal to its concentration, but methaemoglobin absorbs light equally at both 660nm and 940nm so the saturation tends towards a value of 85%
In pulse oximetry:
Oxyhaemoglobin absorbs better at the longer wavelength
True. Pulse oximetry relies on measuring the relative absorbance at two wavelengths (660nm and 940nm). Reduced haemoglobin has greater absorbance at 660nm and oxyhaemoglobin greater at 940nm. The ratio is the determined and the SpO2 is then calculated
In pulse oximetry:
Fetal haemoglobin gives an inaccurately high reading of saturation
False. Fetal haemoglobin does not affect the accuracy of pulse oximetry
In pulse oximetry:
The pulse oximeter is less accurate at low rather than high saturations
True. The monitor is accurate to within 3% at values of over 70%, but less accurate as the saturation falls
Regarding pulse oximetry:
The arterial a.c (pulsatile) component of the signal is approximately 20% of the total absorption
False. The pulsatile component is typically 2% of the total absorption.
Regarding pulse oximetry:
Functional saturation is defined as the ratio of oxyhaemoglobin to oxyhaemoglobin plus reduced haemoglobin
True. Functional saturation is the ratio of O2HB to (O2Hb + Hb), it is Fractional saturation that includes other haemoglobins.
Regarding pulse oximetry:
Beer’s law relates the intensity of transmitted light to the path length
False. Beer’s law relates the absorption of transmitted light to the concentration of the substance whereas Lamberts law relates it to path length.
Regarding pulse oximetry:
Infrared light is only absorbed by small molecules with symmetric bonds
False. Infrared light is absorbed by small molecules with asymmetric bonds so that vibration can occur.
Regarding pulse oximetry:
Motion artefacts produces a high a.c. to d.c. signal ratio
True. Motion will produce an increase in a.c. components however this is noise and therefore the signal to noise ratio will be lower
Causes of inaccuracies on pulse oximetry include:
Methaemoglobinaemia
True. MetHb, hypothermia (causing vasoconstriction) and extraneous lighting can all affect pulse oximetry accuracy.
Causes of inaccuracies on pulse oximetry include:
Fetal haemoglobin
False. Fetal haemoglobin does not have a significant effect.
Causes of inaccuracies on pulse oximetry include:
Hypothermia
True. MetHb, hypothermia (causing vasoconstriction) and extraneous lighting can all affect pulse oximetry accuracy.
Causes of inaccuracies on pulse oximetry include:
Blue nail polish
True. Blue nail varnish absorbs red light and can affect readings. Similarly the use of blue dyes for sentinel node biopsy can lead to reduction in recorded oxygen saturation using a pulse oximeter
Causes of inaccuracies on pulse oximetry include:
Extraneous lighting
True. MetHb, hypothermia (causing vasoconstriction) and extraneous lighting can all affect pulse oximetry accuracy.
Concerning the measurement of oxygen:
A Clark electrode requires a polarising voltage
True. The Clark electrode is a polarographic electrode and needs a polarising voltage.
Concerning the measurement of oxygen:
Fuel cells can be affected by the presence of nitrous oxide
True. Some fuel cells are affected by N2O.
Concerning the measurement of oxygen:
A paramagnetic analyser has a slower response time than a Clark electrode
False. Modern paramagnetic analysers have a fast response time.
Concerning the measurement of oxygen:
A polarographic technique is used in normal blood gas analysers
True. Oxygen is measured by a polarographic technique in most blood gas analysers.
Concerning the measurement of oxygen:
Oxygen saturation is directly measured using a co-oximeter
False. A CO-oximeter measures the concentrations of different haemoglobins and calculates the oxygen saturation from the relevant percentages.
The following are directly measured in a standard blood gas analyser:
pH
False!!? A blood gas analyser directly measures hydrogen ion concentration using a glass electrode. pH is calculated mathematically as the negative Log of the hydrogen ion concentration and therefore not directly measured! Small but important point if asked in the exam.
The following are directly measured in a standard blood gas analyser:
Bicarbonate concentration
A blood gas analyser directly measures hydrogen ion concentration using a glass electrode, the tension of oxygen using a polarographic electrode and carbon dioxide indirectly through a modification of the glass electrode. Bicarbonate is extrapolated using the HH equation.
The following are directly measured in a standard blood gas analyser:
Oxygen tension
True. A blood gas analyser directly measures hydrogen ion concentration using a glass electrode, the tension of oxygen using a polarographic electrode and carbon dioxide indirectly through a modification of the glass electrode.
The following are directly measured in a standard blood gas analyser:
Partial pressure of carbon dioxide
False. A blood gas analyser directly measures hydrogen ion concentration using a glass electrode, the tension of oxygen using a polarographic electrode, and carbon dioxide indirectly through a modification of the glass electrode. Partial pressures are derived by measuring the atmospheric pressure.