Primary FRCA Course Measurement and Monitoring Exam Prep Questions Flashcards
In pulse oximetry:
The SpO2 is determined by the absorbance of light of wavelengths 660nm and 940nm
True. Pulse oximetry relies on measuring the relative absorbance at two wavelengths (660nm and 940nm). Reduced haemoglobin has greater absorbance at 660nm and oxyhaemoglobin greater at 940nm. The ratio is the determined and the SpO2 is then calculated
In pulse oximetry:
Methaemoglobinaemia causes the SpO2 to approach 85%
True. Carboxyhaemoglobin causes an increase in SpO2 equal to its concentration, but methaemoglobin absorbs light equally at both 660nm and 940nm so the saturation tends towards a value of 85%
In pulse oximetry:
Oxyhaemoglobin absorbs better at the longer wavelength
True. Pulse oximetry relies on measuring the relative absorbance at two wavelengths (660nm and 940nm). Reduced haemoglobin has greater absorbance at 660nm and oxyhaemoglobin greater at 940nm. The ratio is the determined and the SpO2 is then calculated
In pulse oximetry:
Fetal haemoglobin gives an inaccurately high reading of saturation
False. Fetal haemoglobin does not affect the accuracy of pulse oximetry
In pulse oximetry:
The pulse oximeter is less accurate at low rather than high saturations
True. The monitor is accurate to within 3% at values of over 70%, but less accurate as the saturation falls
Regarding pulse oximetry:
The arterial a.c (pulsatile) component of the signal is approximately 20% of the total absorption
False. The pulsatile component is typically 2% of the total absorption.
Regarding pulse oximetry:
Functional saturation is defined as the ratio of oxyhaemoglobin to oxyhaemoglobin plus reduced haemoglobin
True. Functional saturation is the ratio of O2HB to (O2Hb + Hb), it is Fractional saturation that includes other haemoglobins.
Regarding pulse oximetry:
Beer’s law relates the intensity of transmitted light to the path length
False. Beer’s law relates the absorption of transmitted light to the concentration of the substance whereas Lamberts law relates it to path length.
Regarding pulse oximetry:
Infrared light is only absorbed by small molecules with symmetric bonds
False. Infrared light is absorbed by small molecules with asymmetric bonds so that vibration can occur.
Regarding pulse oximetry:
Motion artefacts produces a high a.c. to d.c. signal ratio
True. Motion will produce an increase in a.c. components however this is noise and therefore the signal to noise ratio will be lower
Causes of inaccuracies on pulse oximetry include:
Methaemoglobinaemia
True. MetHb, hypothermia (causing vasoconstriction) and extraneous lighting can all affect pulse oximetry accuracy.
Causes of inaccuracies on pulse oximetry include:
Fetal haemoglobin
False. Fetal haemoglobin does not have a significant effect.
Causes of inaccuracies on pulse oximetry include:
Hypothermia
True. MetHb, hypothermia (causing vasoconstriction) and extraneous lighting can all affect pulse oximetry accuracy.
Causes of inaccuracies on pulse oximetry include:
Blue nail polish
True. Blue nail varnish absorbs red light and can affect readings. Similarly the use of blue dyes for sentinel node biopsy can lead to reduction in recorded oxygen saturation using a pulse oximeter
Causes of inaccuracies on pulse oximetry include:
Extraneous lighting
True. MetHb, hypothermia (causing vasoconstriction) and extraneous lighting can all affect pulse oximetry accuracy.
Concerning the measurement of oxygen:
A Clark electrode requires a polarising voltage
True. The Clark electrode is a polarographic electrode and needs a polarising voltage.
Concerning the measurement of oxygen:
Fuel cells can be affected by the presence of nitrous oxide
True. Some fuel cells are affected by N2O.
Concerning the measurement of oxygen:
A paramagnetic analyser has a slower response time than a Clark electrode
False. Modern paramagnetic analysers have a fast response time.
Concerning the measurement of oxygen:
A polarographic technique is used in normal blood gas analysers
True. Oxygen is measured by a polarographic technique in most blood gas analysers.
Concerning the measurement of oxygen:
Oxygen saturation is directly measured using a co-oximeter
False. A CO-oximeter measures the concentrations of different haemoglobins and calculates the oxygen saturation from the relevant percentages.
The following are directly measured in a standard blood gas analyser:
pH
False!!? A blood gas analyser directly measures hydrogen ion concentration using a glass electrode. pH is calculated mathematically as the negative Log of the hydrogen ion concentration and therefore not directly measured! Small but important point if asked in the exam.
The following are directly measured in a standard blood gas analyser:
Bicarbonate concentration
A blood gas analyser directly measures hydrogen ion concentration using a glass electrode, the tension of oxygen using a polarographic electrode and carbon dioxide indirectly through a modification of the glass electrode. Bicarbonate is extrapolated using the HH equation.
The following are directly measured in a standard blood gas analyser:
Oxygen tension
True. A blood gas analyser directly measures hydrogen ion concentration using a glass electrode, the tension of oxygen using a polarographic electrode and carbon dioxide indirectly through a modification of the glass electrode.
The following are directly measured in a standard blood gas analyser:
Partial pressure of carbon dioxide
False. A blood gas analyser directly measures hydrogen ion concentration using a glass electrode, the tension of oxygen using a polarographic electrode, and carbon dioxide indirectly through a modification of the glass electrode. Partial pressures are derived by measuring the atmospheric pressure.
The following are directly measured in a standard blood gas analyser:
Oxygen saturation
False. A blood gas analyser directly measures hydrogen ion concentration using a glass electrode, the tension of oxygen using a polarographic electrode, and carbon dioxide indirectly through a modification of the glass electrode. Oxygen saturation may be derived but is not directly measured
Arterial blood gas samples:
Storage at room temperature causes a rise in pH
False. Blood gas samples become acidotic with time so the pH falls.
Arterial blood gas samples:
A hypothermic patient will have an inaccurately high pO2 if temperature-correction is not employed
True. If temperature-correction is not applied then hypothermic patients will have an inaccurately high PO2 reading.
Arterial blood gas samples:
Air bubbles in the sample may lead to a fall in the pCO2
True. Air bubbles contain no CO2 so will reduce the reading of pCO2.
Arterial blood gas samples:
Excessive heparin will cause an inaccurately high pH
False. Heparin is acidic so will reduce the pH.
Arterial blood gas samples:
Storage at room temperature causes a fall in pO2
True. At room temperature white cells metabolise oxygen and pO2 falls.
The oxygen electrode in a blood gas analyser:
Incorporates on a lead anode
False. The anode is typically silver and the cathode platinum, the electrolyte solution is a potassium chloride solution.
The oxygen electrode in a blood gas analyser:
The cathode is usually platinum
True. The anode is typically silver and the cathode platinum, the electrolyte solution is a potassium chloride solution.
The oxygen electrode in a blood gas analyser:
Potassium hydroxide is the usual electrolyte in the electrode
False. The anode is typically silver and the cathode platinum, the electrolyte solution is a potassium chloride solution.
The oxygen electrode in a blood gas analyser:
Current flow depends on oxygen tension at the cathode
True. Current flow depends on the oxygen tension at the cathode. The more oxygen there is the greater the current (this is the principle of an amperometric sensor). At the cathode a reductive reaction occurs and oxygen is consumed with the gain of electrons to form hydroxide ions.
The oxygen electrode in a blood gas analyser:
Is of the fuel cell type
False. Fuel cells are not used in blood gas analysers. The polarographic (Clark) Electrode is used in blood gas analysers.
The fuel cell:
Contains a lead anode
True. A fuel cell contain a lead anode and gold mesh cathode.
The fuel cell:
Has a faster response time than a polarographic electrode
False. They both have relatively slow response time as it is they are electrochemical processes.
The fuel cell:
Is not affected by temperature
False. It is affected by temperature as a chemical reaction takes place.
The fuel cell:
Nitrous oxide in the gas mixture reacts with the anode to produce nitrogen
True. Nitrous oxide is broken down to nitrogen at the anode.
The fuel cell:
Is used to measure oxygen tension in a blood gas analyser
False. Typically a Clark polarographic electrode is used in a blood gas analyser.