Pilbeams Ch 3 Flashcards
Equation of motion
Pmus + Pvent= V/C+(Raw×flow)
The primary variable that the ventilator adjust to produce inspiration is the control variable
The most commonly used control variables are pressure and volume
Pressure controller
The ventilator maintains the same pressure waveform at the mouth regardless of changes in lung characteristics
Volume controller
Ventilator volume delivery and volume waveform remain constant and are not affected by changes in lung characteristics. Volume is measured
Flow controller
Ventilator volume delivery and flow waveform remain constant and are not affected by changes in lung characteristics. Flow is measured
Time controller
Pressure, volume, and flow curves can change as lung characteristics change. Time remains constant
Phase variables
A phase variable begins, sustains, and, and determines the characteristics of the excretory portion of each breath. 4 phase variables are typically described:
- The trigger variable begins inspiration
- The limit variable limits the pressure, volume, flow, or time during inspiration but it does not end the breath
- The cycle variable ends inspiratory phase and begins exhalation
- The Baseline variable is the end expiratory Baseline (usually pressure) before a breath is triggered
Ventilator determination of actual breath delivery during assisted ventilation
If a patient occasionally starts a breath independently, the ventilator must determine how long to wait before another breath is needed. An example, the rate is set for 6 breaths/minute. The ventilator determines that it has 10 seconds for each breath. If the patient triggers a breath, the ventilator “resets” itself so that it still allows a full 10 seconds (60 s/6 breats) after the start of the patients last breath before it time triggers another breath.