Physics 2 test 2 Flashcards
Quality Control is referring to
Equiptment
Quality Asurance is referring to
Patients
These are consistent with what "Quality" People oriented Necessary and appropriate procedures Producing desired information Accurately interpreted with ALARA as the goal for exposure, cost and inconvenience: scheduling prep instructions report accuracy & distribution image interpretation
Quality Assurance
Began as a war against unacceptable repeat rates
Screen-Film Quality Control
Equiptment oriented-image production, procesing, image evaluation & critique
Screen-Film Quality Control
Planned, Continuous, Documented, Deomstrate adaptations
Screen-Film Quality Control
Best possible image obtained with respect for patient safety
Screen-Film Quality Control
Sort rejects on positioning, motion, density, etc.
Reject analysis (QC)
Goal for reject analysis:
Less than 2% per 250 patients
Who does quality control testing?
Physicist or designated technologist
When and who does Acceptance testing?
Upon installation and it’s done independently
PM
Preventative Maintenance
Routine Performance Evaluation:
Periodic or upon major repair/part replacement
Another reasons why Quality Control testing would be needed
Error Correction Consistency Reproducibility Predictability Confidence
3 benchmarks for QC Guidelines
Tolerance
method
Frequency
QC guidelines for Filtration
2.5 mm Al minimum
Tested by HVL
Annually
QC guidelines for Collimation
Light field coincides withing 2% of SID
Test each size film
Semiannually
QC guidelines for Effective Focal Spot
Tested with slit camera, pinhole camera or star pattern
Variance fairly large
Annually
QC guidelines for kVp
+/- 10 kVp diagnostic (+/- 1 kVp mammo)
Dx: 3% kVp variance will alter image density, radiographic contrast, and patient exposure
Annually
QC guidelines for Exposure Control
Time has a big influence on patient exposure & radiographic density +/- 5%
Test actual time set and AEC systems
backup timer must stop exposure at 6 sec or 600 mAs
Annually
QC guidelines for Linearity
Constant output for any mA/sec resulting constant mAs +/- 10% (reciprocity law)
Measure mR/mAs by varying mA (so timer inaccuracy doesnt flaw test) *only want one variable
Annually
QC guidelines for Reproducibility
Appropriate denstiy & contrast for technique factors selected +/- 5%
half value layer
pieces of equiptment: step wedge, etc
QC guidelines for Film contact
Wire mesh test
semi or annually
QC guidelines for Scren film cleaning
Depends on volume (how many x-rays your performing)
Actually emtpy the cassets and clean them
QC guidelines for Lead protective apparel
45-50 kVp range (not too much go right through-looking for a tear or whole in apron-will look black)
Annually
QC guidelines for Film Illuminators
view box-light is measured in Lumens
Measures averagle light intensity
Annually
Special QC for Fluro
Exposure Rate
Cant exceed (but shouldn't need it anyway) 10 rad/min for ABS 5 rad/min for manual
Special QC for Fluro
Spots
Entrance skin does highest for cassette loaded spots- spots off image intensifier much less exposure smaller mode (mag) image inensifier is highest patient dose *Patient exposure increases from fluro to still spots
Special QC for Fluro
AEC
Constant image brightnes
Measured by seeing if input phosphor receives constant dose regardless of whats being penetrated
Conventional Tomography QC
Tomo
Constant patient exposure
Cut height accurate +/- 5 mm
Patient movement between slices accurate +/- 2 mm
Computed Tomography QC
CT
Ongoing Preventive Maintence Periodic testing of noise/uniformity Linearity Contrast resolution Slice thickness Table incrementation Light localization Patient dose
`Processor QC-Radiographic film processor
DOCUMENTATION: No Doc, it didn’t happen
Digital QC
SMPTE pattern adopted by ACR for evaluation of digital monitors
DICOM Digital Imagning Communication
Line pairs per millimeter
blank spot and line together is a “line pair”