P - Anesthesia Machine Flashcards
Your oxygen supply has failed and you have switched to the backup tanks. Which of the following would be an appropriate measure to reduce the amount of oxygen consumed and prolong the duration of your backup oxygen supply?
Ventilate the patient using the reservoir bag
Most anesthesia machines utilize oxygen as the driving gas to power the ventilator. By reducing fresh oxygen flow rates and eliminating the use of the ventilator by allowing the patient to breathe spontaneously or ventilating via the reservoir bag, you will prolong the backup oxygen supply.
The unidirectional valves in the anesthesia breathing system must be positioned between the patient and the
reservoir bag
Unidirectional valves are typically disc valves or rubber flap valves built into the anesthesia machine breathing circuit. A circle system anesthesia circuit contains two of them: an inspiratory valve and an expiratory valve. The unidirectional valves can be mounted anywhere in the inspiratory and expiratory limbs as long as each valve is positioned between the patient and the reservoir bag on each limb.
You are preparing your anesthesia machine for cases in the morning. You close the pop-off valve, occlude the y-piece, and evaluate for a drop in pressure after pressurizing the circuit to 30 cm H2O using the flush valve. What test have you just performed?
The leak test
The leak test is performed by closing the pop-off valve, occluding the y-piece, and evaluating for a drop in pressure after pressurizing the circuit to 30 cm H2O using the flush valve.
Which statement reflects an accurate understanding of the flowmeter tubes on an anesthesia machine?
A ball is less likely to stick in the presence of dirt
The ball or bobbin in a flowmeter is more likely to stick at low flows where the tube is more narrow. At low flows or in the presence of dirt, a ball is best because it is less likely to stick. The rotameter tubes must be vertical so that the ball or bobbin does not touch the sides of the tube. Flow should be read at the top of a bobbin or in the middle of a ball.
The anesthesia machine fail-safe mechanism will
interrupt nitrous oxide flow if the oxygen supply pressure drops
The oxygen fail-safe device will trigger an alarm and shut off the supply of nitrous oxide in the event of loss of oxygen supply pressure.
The nitrous oxide shutoff device on an anesthesia machine
turns off the nitrous oxide supply when the oxygen pressure drops
The nitrous oxide shutoff device (also known as the fail safe device) triggers an alarm and closes the supply of nitrous oxide and other gases when the oxygen pressure falls below 30 psi.
The hanger-yoke assembly on the anesthesia machine is responsible for which of the following?
Providing uni-directional gas flow
The hanger yoke assembly orients the gas cylinder, provides for uni-directional gas flow, and provides an air-tight seal.
The oxygen proportioning system on an anesthesia machine limits the FiO2 when using nitrous oxide to a minimum of
25%
Anesthesia machines are configured to prevent the delivery of a hypoxic mixture of gases from the flowmeters. One of the ways this is done is to link the nitrous oxide flowmeter to the oxygen flowmeter so that at least 25% oxygen is delivered when nitrous oxide is being used.
You are using the nitrous oxide tank connected to your anesthesia machine. Your flowmeters are set to administer 1 liter per minute of oxygen and 2 liters per minute of nitrous oxide. The nitrous oxide tank just dropped below a pressure of 745 psig. About how many more minutes of nitrous oxide do you have left?
200 minutes
Because E-cylinder nitrous oxide tanks contain nitrous oxide in both the liquid and gas state, the only accurate way to determine the amount of gas left in the tank is by weighing it. A full E-cylinder tank will hold 1590 liters of nitrous oxide. When the pressure gauge drops below 745 psig, the liquid form has been consumed. The amount of nitrous oxide in the gas phase at that point is about a quarter of the tank or around 400 liters. At 2 liters per minute, you would have approximately 200 minutes of nitrous oxide left.
All of the following are tasks that oxygen performs in the anesthesia machine except
pressurize the variable-bypass vaporizers to 2 atmospheres
Unlike other medical gases, oxygen serves multiple functions in the anesthesia machine. The five tasks of oxygen in the anesthesia machine are: supply the fresh gas flowmeter, power the oxygen flush valve, activate the oxygen low-pressure alarms, activate the fail-safe mechanism, and drive the ventilator bellows. Oxygen passes through the variable-bypass vaporizers when they are turned on, but it does not increase the pressure within them.
Match the anesthesia machine component with the pressure system to which it belongs.
hangar yoke, flowmeters, O2 flush valve ——high, intermediate and low pressure system
hangar yoke - high
O2 flush valve- intermediate
flowmeters - low pressure
The anesthesia machine is divided into three pressure systems: the high-pressure system, intermediate-pressure system, and the low-pressure system.
When the BAG/VENT selector on the anesthesia machine is set to VENT, which components are eliminated from the circle anesthetic system? (select two)
APL valve & Gas reservoir bag
When the BAG/VENT selector is switched to VENT, controlled ventilation is instituted and the gas reservoir bag and APL valve are eliminated from the circle system.
On what principle do the mechanical flowmeters on anesthesia machines function?
Flow past a resistance is proportional to the pressure
Mechanical flowmeters function on the principle that flow past a resistance is proportion to the pressure. The flowmeter actually measures the drop in pressure that occurs when the gas flow passes through a resistance.
Which pin configuration in the pin index safety system is used by oxygen?
2&5
The pin index safety system utilizes two 5 mm pins that have to align with the matching connector. This is to prevent the inadvertent connection of a wrong gas cylinder to an anesthesia machine. Each specific gas has its own pin configuration. For air, it is pins 1 and 5, for nitrous oxide, it is pins 3 and 5, and for oxygen, it is pins 2 and 5.
What are the two components of the anesthesia machine circle system test? (select two)
the flow test
The leak test
The circle system test evaluates the circle breathing system from the common gas outlet to the y-piece and consists of two parts, the leak test and the flow test.
You are testing your anesthesia machine in the morning before cases begin. You disconnect the pipeline supply and turn on the backup oxygen cylinder to make sure it functions correctly. The oxygen cylinder is full and connected correctly, but the anesthesia machine still does not pressurize. You know that the problem could be due to a
fault in the oxygen pressure failure device
If you have verified that the backup oxygen cylinder is full, connected properly, and opened, but the anesthesia machine still does not pressurize, then there is a problem in the high or intermediate pressure system in the anesthesia machine. The flowmeters, vaporizers, and unidirectional valves are all part of the low pressure system.
Which component is not part of the low pressure system on an anesthesia machine?
Second-stage pressure regulator
The low pressure system on an anesthesia machine contains the flowmeters, hypoxia prevention devices, unidirectional valves, pressure relief devices, and the common gas outlet. The intermediate pressure system of an anesthesia machine contains the pneumatic part of the master switch, pipeline inlet connections and pressure indicators, the gas power outlet, oxygen flush valve, oxygen pressure failure devices, and flow control valves.
Which of the following are components of the low pressure system of an anesthesia machine? (select four)
flowmeters, unidirectional valves, common gas outlet, vaporizers
The low pressure system on an anesthesia machine contains the flowmeters, vaporizers, hypoxia prevention devices, unidirectional valves, pressure relief devices, and the common gas outlet. The intermediate pressure system of an anesthesia machine contains the pneumatic part of the master switch, pipeline inlet connections and pressure indicators, the gas power outlet, oxygen flush valve, oxygen pressure failure devices, and flow control valves.
The function of the vaporizer interlock system on an anesthesia machine is to
allow only one vaporizer to be turned on at a time
The function of the vaporizer interlock system on an anesthesia machine is to allow only one vaporizer to be turned on at a time.
On an anesthesia machine, what components are found in the intermediate pressure system? (select four)
pipeline inlet connections
O2 flush valve
flow control valves
pipeline pressure indicators
The intermediate pressure system of an anesthesia machine contains the pneumatic part of the master switch, pipeline inlet connections and pressure indicators, the gas power outlet, oxygen flush valve, oxygen pressure failure devices, and flow control valves.
Select two components of the anesthesia machine that are in the intermediate pressure system.
Oxygen supply failure alarm
Oxygen flush valve
The intermediate pressure system of an anesthesia machine contains the pneumatic part of the master switch, pipeline inlet connections and pressure indicators, the gas power outlet, oxygen flush valve, oxygen pressure failure devices, and flow control valves.
The device that connects the oxygen cylinder to the anesthesia machine is known as the
hanger yoke
Gas cylinders are connected to the back of the anesthesia machine by the hanger yoke. It has a safety mechanism known as the pin index safety system that is designed to prevent the inadvertent connection of the wrong gas to the yoke.
The low-pressure circuit leak test would not assess the integrity of the
cylinder pressure indicator
The low-pressure circuit leak test evaluates the anesthesia machine from the flow control valves to the common gas outlet. This includes the portion of the machine that is downstream from all safety devices except the oxygen analyzer. The cylinder pressure indicator is in the intermediate pressure system.