Novi Flashcards
What does ERI stand for ?
Elective replacement indicator
What does EOS stand for?
End of Service
What does EoBL stand for?
End of battery life
What does the first warning indicate on the novi ipg?
Stimulation will soon not be available (4weeks)
What does the second warning indicate on the novi ipg?
Stimulation is NOT available
What is the implant depth for precision novi?
No depth!
What two ports are on Novi? Why?
C and D, Primary cell IPG only requires the primary chip
What is the ETS/IPG range for the wand to connect it to the CP? How close must the wand get to the ipg to connect? Why is this important?
22inches, 18 inches, because when interop testing we don’t want to break the steril field.
What leads are compatible with Novi?
Infinion trial, Infinion CX, Linear ST and Linear 3-4, artisan 2x8, AVISTA (no MRI conditionality)
What is the minimum time between ERI and EoBL?
4 weeks
Can you program when ERI is enabled?
No
End of Service means stimulation is NO longer available. T or F.
True
Can you connect your ipg to the remote if EoBL indicator is present?
No
What does EUI stand for?
Energy Use Index
What does the battery estimate show us from the program options?
Recharge estimate and EUI