MRI MODE Flashcards
What ipg and remote system is MRI mode available on?
WWA
What does the system check before entering into MRI Mode? 3 things
Lead impedance
Lead check
Ipg battery level
Once MRI mode is enabled, what two things happen?
Stim turns off and IPG pumps high voltage to protect system electronics
In order to turn therapy back On after a MRI Scan, what must happen?
We must exit out of MRI mode
During the systems checks in MRI mode, what 3 error screens could display?
- Stimulator battery is low
- impedance check detects and anomaly
- there is an error in the stimulator
True or false: primary cell IPG’s can still be placed in MRI mode if they are at ERI or EOS.
False
T or F: an ipg battery must be fully charged for a patient to undergo MRI because it protects the patient from unwanted stim and it needs to pump high voltage to protect the system.
True
What is the active scan time limit for WWA?
30 mins
T or F patients with any pre-existing BSCI lead (50 or 70 that have a battery swap to a WWA will maintain full body mri conditionality even if they have adapters, splitters or extensions implanted
FALSE
What leads are MRI conditional with WWA system?
ALL 50 & 70 CM PERC AND PADDLE LEADS
What must be done before a patient can undergo an MRI with WWA?
IPG fully charged and MRI mode successful enabled
T or F: All implanted NEUROMOD devices are MR conditional including competitors?
True
T or F: Assuming all conditions have been met, a patient must undergo a scan using B1rms regardless of the Alpha Device they have implanted.
TRUE