MOSBY ONLINE - TREATMENT PLANNING Flashcards
Tissue compensators are used to:
A. provide uniform dose distribution in tissue deficit areas
B. delineate field arrangements in soft tissue
C. register patient position during treatment
D. shape the treatment field around the tumor
A. provide uniform dose distribution in tissue deficit areas
Rationale:
Tissue compensators are used in areas where there are tissue deficits. The compensator helps provide uniform dose distribution in this area.
A dose of 300 cGy was to be delivered to the thoracic spine at 95 cm SSD. An error caused treatment to be given at an SSD of 92 cm. The actual dose delivered was:
A. 281
B. 291
C. 320
D. 341
C. 320
Rationale:
The inverse square law applies here. 300(95/92)subscript2 = 320.
While treating lateral fields in the treatment of the craniospinal axis, to avoid having the caudal margin of the lateral fields diverge into the posterior spine field, the couch should be rotated:
A. way from the collimator
B. toward the collimator
C. until the table is at 5 degrees
D. until the table is at 10 degrees
B. toward the collimator
Rationale:
To prevent divergence of the lateral fields into the posterior field, the couch should be rotated in toward the collimator and degrees calculated by considering the field size and distance.
Perforated thermoplastics are better than nonperforated thermoplastics because they:
A. reduce the average skin dose
B. increase the average skin dose
C. are much more sturdy D. make it easy to reproduce the setup
A. reduce the average skin dose
Rationale:
Perforated thermoplastics reduce the average skin dose compared to nonperforated thermoplastics.
When using lateral and anterior supraclavicular fields to treat a tumor in the oropharynx, the field junction should be placed:
A. cephalad of the thyroid notch
B. caudal of the thyroid notch
C. caudal of the larynx
D. cephalad of the mandible
A. cephalad of the thyroid notch
Rationale:
If field matching is done while treating the oropharynx, the junction can reasonably be placed cephalad, or superior to, the thyroid notch to avoid overdosing the larynx.
“cephalad” means what?
superior
“caudad” means what?
inferior
Increasing the field size from 5 x 5 to 20 x 20, with all other factors remaining the same, the percent depth dose will:
A. increase
B. decrease
C. remain the same
D. move the Dmax point
A. increase
Rationale:
Increasing the field size will increase the percent depth dose.
FS and PDD are directly related!!!!!
The volume of lung within an anterior supraclavicular field can be reduced when treating breast carcinoma by:
A. elevating the arm
B. treating the supraclavicular field every other day
C. placing the angle board under the chest
D. moving the breast down on the chest wall and making the match line as superior as possible
D. moving the breast down on the chest wall and making the match line as superior as possible
Rationale:
To keep the volume of lung as low as possible, the match line should be as superior as possible. This may be achieved in women with larger breasts by using a position or aid that will shift the breast tissue down.
The tolerance dose (TD 5/5) for the whole brain is:
A. 45 Gy in 25 fractions
B. 60 Gy in 30 fractions
C. 60 Gy in 15 fractions
D. 45 Gy in 15 fractions
A. 45 Gy in 25 fractions
Rationale:
The tolerance dose for the whole brain is 45 Gy in fractions of 1.8 to 2.0 Gy.
Where is the depth of the 80% isodose line for 12 MeV electron beam through water?
A. 3 cm B. 4 cm C. 6 cm D. 10 cm
B. 4 cm
Rationale:
Use the 80% rule. 1/3 of the electron energy is the approximate depth in centimeters of the 80% isodose line.
An excerpt from an equivalent square table shows: Using the table, the equivalent square for a field size 9.5 x 13 would be:
A. 10.6 B. 10.9 C. 11.3 D. 11.45
B. 10.9
Rationale:
Using interpolation between the values 10.6 and 11.3. Take the average to get 10.95.
The closest blocked equivalent square for a 10 x 20 rectangular field with a 4 x 6 block in the field is:
A. 8.6
B. 11.4
C. 13.3
D. 15.6
C. 13.3
Rationale:
Subtract the area of the block from the area of the open field, and then take the square root. 10 x 20 = 200, 4 x 6 = 24. 200-24 = 176. The square root of 176 = 13.3.
Bolus should be composed of a material:
A. having higher absorption than tissue
B. having lower absorption than tissue
C. easy to clean with disinfectant solutions
D. equivalent to tissue in absorption and scattering properties
D. equivalent to tissue in absorption and scattering properties
Rationale:
Bolus should be made of tissue-equivalent material.
The dose under a 5 HVL block would be approximately:
A. 3% to 5% of the given dose
B. 5% to 10% of the given dose
C. 10% to 20% of the given dose
D. 20% to 50% of the given dose
A. 3% to 5% of the given dose
Rationale:
A block 5 half-value layers thick should allow about 3% of the dose applied.
When a missing tissue compensator is used, it should be placed:
A. directly on the skin surface
B. at some distance from the skin surface
C. on the end of an electron cone
D. on top of a bolus material at isocenter
B. at some distance from the skin surface
Rationale:
To preserve the skin-sparing effect, tissue compensators should be placed at least 15 cm from the patient’s skin surface.
The field size of an X-ray or gamma ray beam is defined as the area enclosed within the boundary of the __________% depth dose line.
A. 50 B. 60 C. 80 D. 100
A. 50
Rationale:
The 50% isodose line defines the field size of X-ray or gamma ray beams.
When using rotational arc therapy, monitor units or treatment time should be calculated using which of the following methods?
A. Mayneord's factor B. Paterson-Parker C. tissue air ratio D. Clarkson's
C. tissue air ratio
Rationale:
Rotational arc treatment time/monitor units are calculated using the TAR method.
Field gaps or abutting junctions are often shifted (feathered) a few times during the course or treatment in order to:
A. reduce the risk of nausea and vomiting
B. reduce the risk of overdosing critical structures
C. make the gap larger
D. change the exit point
B. reduce the risk of overdosing critical structures
Rationale:
Feathering matching junctions, or gaps, are intended to decrease the risk of overdosing or underdosing critical structures.
Total dose to the lumpectomy site following accelerated partial breast irradiation utilizing high dose rate brachytherapy will be approximately:
A. 20 Gy B. 35 Gy C. 45 Gy D. 60 Gy
C. 45 Gy
Rationale:
Using APBI techniques, dose to the lumpectomy site typically ranges from 40-50 Gy
Posterior cervical nodes need to be treated and the patient’s shoulder will only allow a treatment distance of 115 cm SSD on the 100 cm SAD linear accelerator. The inverse square factor needed in the monitor unit calculation would be:
A. (100/115)2 B. (115/100)2 C. (100/15)2 D. (15/100)2
A. (100/115)2
Rationale:
The inverse square factor in extended distance dose calculations is computed by the relationship: (SAD/SSD + Dmax)superscript2.
Since the energy is not specified here, Dmax can not be accounted for in this exercise.
The attenuation of Cerrobend is less dense than lead by about:
A. 5%
B. 10%
C. 15%
D. 25%
C. 15%
Rationale:
Cerrobend or Lipowitz metal is about 15% less dense than lead.
Cerrobend or Lipowitz metal is about ____% less dense than lead.
15%
The tumor lethal dose of 50 cGy will need to be given to a midline abdominal tumor through a single anterior port using the SSD technique. The abdominal tumor is located at the depth of the 80% dose line, and the spinal cord is at the depth of the 75% dose line. Dose to the cord will be:
A. 66 Gy B. 50 Gy C. 47 Gy D. 37.5 Gy
C. 47 Gy
Rationale:
The spinal cord will receive 47 Gy.
You can compute the applied dose by using TD/%DD. 50/.80 = 62.5 Gy and then taking 75% of the applied dose- 62.5(0.75) = 47 Gy.
Or use direct proportion; 50 Gy:80% as x:75%.
The focus-to-film distance on a conventional simulator is 140 cm. A 5 cm wire is placed on the patient’s skin. The focus-to-skin distance is 100 cm. The magnification factor is given by:
A. 100/5
B. 140/100
C. 100/140
D. 5/100
B. 140/100
Rationale:
The magnification formula is FFD/FOD = magnification. Therefore, 140/100 = 1.4.
A treatment designed to be given at 120 cm SSD is mistakenly given at 100 cm SSD. What is the error in dose delivered?
A. 20% overdose B. 20% underdose C. 44% overdose D. 44% underdose
C. 44% overdose
Rationale:
Use the inverse square theory. The distance has changed by 20%, so the dose must change in the opposite direction by a square of the distance changed. Check your work by substituting a number for an initial intensity.
A dose of 55 Gy will be given through anterior and lateral ports to the maxillary sinus. The lens is included in the treatment fields. The therapeutic ratio is approximately:
A. 0.9 B. 0.18 C. 5.5 D. 11
B. 0.18
Rationale:
The tolerance dose of the lens ranges from 5 Gy to 10 Gy depending on the source. Use the upper limit of the range and the therapeutic ratio formula: 1000/5500 = 0.18.
The purpose of an immobilization device is to:
A. make the patient’s position comfortable
B. facilitate reproducing and maintaining the position
C. prevent the patient from falling off the PSA
D. help the therapist set up the treatment faster
B. facilitate reproducing and maintaining the position
Rationale:
Immobilization devices should help in reproducing position and maintaining position.
Dose rates in air express the dose measured:
A. at 10 cm depth in a phantom
B. without a phantom, with a buildup cap at a certain distance
C. with a phantom, without a buildup cap at a certain distance
D. without a build up cap or phantom with a 10 x 10 cm field size
B. without a phantom, with a buildup cap at a certain distance
Rationale:
Dose rates in air are measured without a phantom and with a buildup cap to achieve electronic equilibrium at a certain distance from the source.
The angle between two beams is known as the :
A. arc angle B. hinge angle C. wedge angle D. divergent angle
B. hinge angle
Rationale:
The degree of separation between two beams is the hinge angle.
If the exposure rate at 1 meter is 53.7 R/minute, at what distance is the exposure rate 35 R/minute?
A. 0.5 meters
B. 0.75 meters
C. 1.53 meters
D. 1.24 meters
D. 1.24 meters
Rationale:
Using the inverse square formula. 53.7/35 = 12 /x2, 1.53 = x2; x = 1.24.
Rotational arc therapy is elected for treatment to the pituitary gland. The arc is 240 degrees and the calculated monitor units are 196 MU. The MU/degree setting will be:
A. 0.817
B. 0.667
C. 1.22
D. 1.5
A. 0.817
Rationale:
To calculate the MU/degree: 196 MU/240 degrees = 0.817 MU/degree.
While using adjacent photon fields for treatment:
A. It doesn’t mater where the hot spot occurs.
B. It doesn’t matter where the cold spot occurs.
C. Dose nonuniformity due to error in abutting can be minimized by increasing the lengths every other day.
D. Dose uniformity can be improved by moving the junction or gap at regular intervals during treatment.
D. Dose uniformity can be improved by moving the junction or gap at regular intervals during treatment.
Rationale:
Adjacent photon fields should have a skin gap, and/or the junction should be feathered periodically.
When preparing to assist with the administration of iodine based contrast, the therapist should inspect the contrast container and use only when the content is:
A. expired
B. clear
C. viscous
D. warm
B. clear
Rationale:
Before use, contrast should be inspected and be confirmed as clear, not discolored or cloudy, and having a current date.
The physical size of the treatment field is defined at the interception of the central axis at the specific isocentric distance and which isodose line?
A. 10%
B. 20%
C. 50%
D. 80%
C. 50%
Rationale:
The geometric field size is defined at isocenter at the interception of the 50% isodose line.
A 6 MV photon beam on the linear accelerator has an output at isocenter of 1.2 cGy/MU. What is the output at an extended distance of 120 cm?
A. 0.8124
B. 0.9876
C. 1.45
D. 1.776
A. 0.8124
Rationale:
The new output can be found by multiplying the old output by the inverse square factor 0.677. Use the inverse square formula.
The tolerance dose (TD 5/5) for the whole lung in the adult is:
A. 1800 cGy in 5 fractions
B. 1500 cGy in 5 fractions
C. 1800 cGy in 10 fractions
D. 2400 cGy in 10 fractions
C. 1800 cGy in 10 fractions
Rationale:
The tolerance dose of the whole lung in the adult is about 17.5 Gy when standard fractionation is used:180 to 200 cGy/fraction.
Indian club needles have:
A. gold seed
B. one hot end
C. two hot ends
D. even activity
B. one hot end
The physician asks you to add a 5 x 4 corner block to a 12 x 14 open lung field. The blocked equivalent square is:
A. 7 x 7
B. 7 x 10
C. 12 x 12
D. 13 x 13
C. 12 x 12
Rationale:
To find the blocked equivalent square, subtract the blocked area from the open area. 12(14)-5(4) = 148. The square root of 148 = 12.1.
Diagnostic CT machines are often not suitable for therapy planning because (select two):
- The table top is concave.
- The bore is not large enough to accommodate various positions and devices.
- The slice thickness cannot be varied.
- The mode can not be switched from axial to helical.
- The table top is concave.
- The bore is not large enough to accommodate various positions and devices.
Rationale:
Diagnostic CT scanners do not have bores large enough to accommodate positions and positioning devices used in therapy. Although the table top is concave, the diagnostic CT table often has an insert to make the table flat.
An advantage to using custom compensators for missing tissue instead of bolus in megavoltage therapy is:
A. reduced penumbra
B. preservation of skin sparing
C. increased surface dose
D. increased exit dose
B. preservation of skin sparing
Rationale:
Custom compensators are placed near the collimators at a substantial distance away from the patient. This preserved the skin sparing effect of the megavoltage beam.
The energy loss of megavoltage (MeV) electron beams in water is approximately_____MeV/cm.
A. 10 B. 5 C. 2 D. 3
C. 2
Rationale:
Electron beams lose about 2 MeV per centimeter of tissue.
One disadvantage of using a vertex field to treat primary brain malignancies is that the field exits through the:
A. brain and mediastinum
B. pharynx and spinal cord
C. orbits
D. chest
B. pharynx and spinal cord
Rationale:
When treating the vertex with the patient in neutral position, the vertex field would exit through the pharynx. If the patient is treated with chin flexed, the vertex field exits through the pharynx and spinal cord.
A patient was to have received 250 cGy each fraction for 10 fractions at a distance of 100 cm SSD. A readout error demonstrated that the patient was actually treated at a distance of 105 cm SSD. What was the actual total dose delivered?
A. 2267 cGy
B. 2381 cGy
C. 2756 cGy
D. 2625 cGy
A. 2267 cGy
Rationale:
Use the inverse square. 2500(100/105)2 = 2267.
The posterior surface of the patient should be flat when treating the entire spine so that:
A. Skin gaps are smaller.
B. Dose is more uniform.
C. Position is reproducible.
D. Patients are most comfortable.
B. Dose is more uniform.
Rationale:
When treating the entire spine, the posterior surface should be as flat as possible so that dose distribution is uniform.
A TBI patient will be treated on a 100 cm isocentric accelerator at a distance of 450 cm. The calibrated dose rate at isocenter is 250 cGy/minute. The dose rate at the treatment distance is:
A. 55.5 cGy/minute
B. 1125 cGy/minute
C. 62.5 cGy/minute
D. 12 cGy/minute
D. 12 cGy/minute
Rationale:
Use the inverse square formula. 250(100/450)2 = 12.
A patient is receiving 3000 cGy total dose to a spine port using the SSD technique. The total given dose is 3636 cGy. What is the percent depth dose?
A. 100% B. 120% C. 82.5% D. 75%
C. 82.5%
Rationale:
Use the applied dose formula. Given dose = TD/%DD. 3636 = 3000/x; 3000/3636 = 0.825.
The CTV is larger than the GTV because it includes the:
A. normal tissue
B. margins for daily uncertainties
C. areas of potential positive disease
D. lymphatic drainage
C. areas of potential positive disease
Rationale:
The CTV includes the tumor volume and areas of potentially positive disease.
The output of a 7 MeV electron beam, 12 x 12 cm field is 1.01 cGy/MU. Calculate the monitor units required to deliver 250 cGy to the depth of the 90% dose line.
A. 275 MU
B. 223 MU
C. 227 MU
D. none of the above
A. 275 MU
Rationale:
Use the monitor unit formula for SSD technique.
The ratio of dose at depth in tissue to the dose at electronic equilibrium on the beam axis is known as the:
A. tissue air ratio
B. percent depth dose
C. scatter air ratio
D. field size correction factor
B. percent depth dose
Rationale:
The percent depth dose is the ratio of dose at Dmax to dose at a certain depth in tissue.
ratio of dose at Dmax to dose at a certain depth in tissue = ______________
PDD
In order to reduce the potential for electron contamination, the preferred material for custom block and compensator holders is:
A. foil
B. Lucite
C. aluminum
D. tungsten
B. Lucite
Rationale:
Lucite is a sturdy plastic and is preferred over the other listed materials to reduce the potential for secondary, particle contamination.
An excerpt from an equivalent square table shows:
Using the table, the equivalent square for a field size 8 x 13.5 would be:
A. 9.3 B. 9.9 C. 10 D. 10.25
C. 10
Rationale:
Using interpolation between the values 9.9 and 10.1, the equivalent square for 8 x 13.5 would be 10 cm.
When treating adjacent, multiple fields on a patient:
A. Patient position should remain the same for all fields.
B. Field borders should abut on the patient’s skin.
C. Fields should be treated on alternate days.
D. Patients should be positioned supine.
A. Patient position should remain the same for all fields.
Rationale:
For multiple, adjacent fields, the patient’s position should remain the same to avoid hot and cold spots.
Calculate the equivalent square for a rectangular field 10 x 15.
A. 12 cm B. 12.5 cm C. 13.5 cm D. 13 cm
A. 12 cm
Rationale:
Use Sterling’s formula.
The clinical target volume (CTV) includes:
A. gross target and planning target volume
B. gross target and irradiation volume
C. gross target and subclinical disease
D. planning target and irradiation target volume
C. gross target and subclinical disease
Rationale:
The CTV is the gross tumor plus subclinical disease.
Total body irradiation must be administered with the patient at an extended distance of 450 cm SSD on our 100 cm SAD linear accelerator. The dose rate at isocenter is 250 cGy/min. The dose rate at the new treatment distance will be:
A. 12.3 cGy/min
B. 810 cGy/min
C. 55.5 cGy/min
D. 506 cGy/min
A. 12.3 cGy/min
Rationale:
The inverse square law applies. As distance changes, the dose rate (intensity) changes inversely, by a square of the distance.
Which of the following radioisotopes is sometimes administered into the peritoneal cavity in the treatment of gynecological malignancies?
A. Cesium-137 B. Iridium-192 C. Iodine-131 D. Phosphorus-32
D. Phosphorus-32
Rationale:
Phosphorus-32 can be suspended in liquid and administered intracavitarily for the management of gynecologic malignancies that have spread to the abdominal cavity.
Rotational arc therapy is elected for treatment to the pituitary gland. The arc is 240 degrees and the calculated monitor units are 196 MU. The MU/degree setting will be:
A. 0.817 B. 0.667 C. 1.22 D. 1.5
A. 0.817
Rationale:
To calculate the MU/degree: 196 MU/240 degrees = 0.817 MU/degree.
Adjacent fields with respective lengths of 40 cm and 36 cm should intersect at a depth of 10 cm. The patient will be treated on a 100 cm isocentric accelerator using the SSD technique. Calculate the skin gap.
A. 4.4 cm
B. 3.4 cm
C. 3.8 cm
D. 2.2 cm
C. 3.8 cm
Rationale:
Gap formula is L/2 (d/SSD) for field one + L/2 (d/SSD) for field two.
A good surface landmark to approximate the apex of the lung is the:
A. clavicle B. sternal angle C. thyroid notch D. cricoid notch
A. clavicle
Rationale:
The apex of the lung is approximately 3 cm superior to the clavicle.
The linear accelerator dose rate is 250 MU/minute. What will be the treatment time for a treatment field receiving 100 MU?
A. 0.4 seconds
B. 2.5 seconds
C. 4.0 seconds
D. 24 seconds
D. 24 seconds
Rationale:
At 250 MU/minute. Using a direct proportion, it will take 0.4 minutes to deliver 100 MU. Convert the 0.4 minutes to 24 seconds also using direct proportion.
Assuming ideal patient contour, two fields utilizing 30 degree wedges have an optimal hinge angle of:
A. 20 degrees
B. 180 degrees
C. 120 degrees
D. 60 degrees
C. 120 degrees
Rationale:
Use the hinge angle formula.
A tumor that can be completely eradicated by radiation is known to be:
A. radioresistant B. radiocurable C. reoxygenated D. recurrent
B. radiocurable
Rationale:
A tumor that can be completely eradicated by radiation is called radiocurable.
Surface mold brachytherapy is appropriate for all of the following except:
A. pelvic sidewall tumors
B. hard palate tumors
C. orbit tumors
D. superficial skin cancer
A. pelvic sidewall tumors
Rationale:
Mold therapy is appropriate in superficially located tumors in the oral cavity, skin, or eye.
The 50% isodose line for a 12 MeV electron beam would be at an approximate depth of:
A. 3 cm
B. 4 cm
C. 5 cm
D. 6 cm
C. 5 cm
Rationale:
Use the rules for electron beam therapy: 4(E) = depth of 50% isodose in millimeters. 4(12) = 48 mm or E = 2.5 x depth of 50% isodose in centimeters. 12 = 2.5(x).
Percent depth dose varies with:
- depth
- fractionation
- field size
- radiation qualityA. 1, 3
B. 2, 4
C. 1, 3, 4
D. 1, 2, 3, 4
C. 1, 3, 4
Rationale:
Percent depth dose varies with the depth in tissue, the field size, and beam energy.
Kidney tolerance (TD 5/5) in the adult would be:
A. 2000 cGy delivered in 5 fractions
B. 1500 cGy delivered in 5 fractions
C. 2000 cGy delivered in 10 fractions
D. 2400 cGy delivered in 10 fractions
C. 2000 cGy delivered in 10 fractions
Rationale:
The tolerance dose of the kidney in the adult is about 1800 to 2300 cGy using standard fractionation of 180 to 200 each fraction.
Calculate the monitor units necessary to deliver 150 cGy to a depth of 5 cm using a single field SAD technique. The reference dose rate is 0.998 cGy/MU. The field size factor is 1.02 and the TAR is .856.
A. 172 MU B. 147 MU C. 176 MU D. 182 MU
A. 172 MU
Rationale:
Use the monitor unit formula. TD/dose rate x field size factor x TAR.
Radiation therapy to treat medullary thyroid cancer requires large treatment portals that include:
1. the lesion
2. bilateral supraclavicular nodes
3. superior mediastinum
A. 1, 2
B. 1, 3
C. 2, 3
D. 1, 2, 3
D. 1, 2, 3
Rationale:
If medullary thyroid cancer is treated with radiation, the field will include the lesion, bilateral supraclavicular nodes, and superior mediastinal nodes.
Because electrons scatter so readily in the air, they are:
A. removed before striking the target
B. focused by cones extending close to isocenter
C. bent by 90 or 270 degrees toward the collimator
D. used only at distances shorter than the SAD
B. focused by cones extending close to isocenter
Rationale:
Cones are used to help keep electrons focused due to their potential to scatter in air.
In the usual orientation of the three planes with respect to the patient:
A. The transverse plane lies across the patient.
B. The sagittal plane lies across the patient.
C. The coronal plane lies parallel with the x axis.
D. The sagittal plane lies parallel with the z axis.
A. The transverse plane lies across the patient.
Rationale:
The transverse plane corresponds to the axial plane and lies across the patient, dividing the patient into upper and lower halves
Using a 100 cm SAD machine, an anterior SSD reading of 86.5 cm, and posterior SSD reading of 85.5 cm are given. The total anterior to posterior separation along the central axis is:
A. 14.5 cm
B. 24 cm
C. 28 cm
D. 14 cm
C. 28 cm
Rationale:
The separation can be found by adding the depths of isocenter in parallel opposed fields. 13.5 + 14.5 = 28.
Hypogastric lymph nodes would likely be included in radiation field for the treatment of the:
A. base of tongue
B. stomach
C. esophagus
D. rectum
D. rectum
Rationale:
The hypogastric lymph nodes are part of external iliac chain of nodes in the pelvis and would likely be included in fields for rectal carcinoma.
An entire spine needs treatment with field dimensions of 9 x 45. The 100 cm SAD linear accelerator to be used has a collimator limit of 40 x 40. What treatment distance will be required?
A. 89 cm SSD B. 140 cm SSD C. 105 cm SSD D. 112.5 cm SSD
D. 112.5 cm SSD
Rationale:
Using direct proportion. Take the largest dimension needed. 4:100 as 45:x. 40(x) = 45(100). X = 112.5.
The dose rate on a linear accelerator is 200 MU/minute. How long is a treatment requiring 240 MU?
A. 0.694 minutes
B. 0.833 minutes
C. 1.2 minutes
D. 1.44 minutes
C. 1.2 minutes
Wedge angle refers to:
A. the angle of the actual wedge filter
B. the angle at which an isodose curve at a specific depth is tilted as a result of the wedge being inserted in the beam
C. the angle at which an isodose curve at a specified depth is tilted with respect to the central axis of the beam
B. the angle at which an isodose curve at a specific depth is tilted as a result of the wedge being inserted in the beam
Rationale:
The tilt of the isodose curve at a specified depth when a wedge filter is inserted is how the wedge angle is specified.
If the output of a machine is 130 rad/min at 80 cm, what is the output at 60 cm?
A. 65 rad/min
B. 231 rad/min
C. 110 rad/min
D. 150 rad/min
B. 231 rad/min
Rationale:
Use the inverse square. 130 (80/60)2 = 231.
Assuming ideal patient contour, a suitable wedge angle for two fields separated by 90 degrees would be:
A. 30 degrees
B. 45 degrees
C. 60 degrees
D. 90 degrees
B. 45 degrees
Rationale:
Use the hinge angle formula. Hinge = 180 - 2(w). 90 = 180 - 2(x).
An entire spine is treated by matching two fields both treated at 100 cm SSD. Field number 1 measures 8 x 20 and field number 2 measures 8 x 25. In order to match these fields at a depth of 5 cm, a_______cm skin gap would be required.
A. 1.1 cm B. 2.0 cm C. 5.0 cm D. 3.2 cm
A. 1.1 cm
Rationale:
Use the gap formula. L/2(d/SSD) + L/2 (d/SSD) = skin gap
20/2 (5/100) + 25/2 (5/100) = 1.1 cm.
What is the equivalent square for a 15 x 6 field?
A. 9.5 x 9.5 B. 8.6 x 8.6 C. 10 x 10 D. 2.1 x 2.1
B. 8.6 x 8.6
Rationale:
Use Sterling’s formula. 2(w x L)/ w + L. 2(15 x 6)/ 15 + 6 = 8.6.
A TBI patient will be treated on a 100 cm isocentric accelerator at a distance of 450 cm. The calibrated dose rate at isocenter is 250 cGy/minute. The dose rate at the treatment distance is:
A. 55.5 cGy/minute B. 1125 cGy/minute C. 62.5 cGy/minute D. 12 cGy/minute
D. 12 cGy/minute
Rationale:
Use the inverse square formula. 250(100/450)2 = 12.
A patient is treated using the SAD technique on a 100 cm isocentric linear accelerator. The field size is 24 x 20. The isocenter is located 5 cm below the patient’s skin surface. What does the field measure on the patient’s skin?
A. 24 x 20
B. 19 x 15
C. 23 x 19
D. 22 x 22
C. 23 x 19
Rationale:
Use direct proportion.
Custom tissue compensators are used to:
A. filter scattered electrons
B. reduce the beam energy
C. compensate for surface irregularity
D. none of the above
C. compensate for surface irregularity
Rationale:
Custom compensators compensate for surface irregularities.
In the maintenance phase of managing ALL, whole brain irradiation may be given with the inferior border of the treatment field at the:
A. base of skull B. mastoid tip C. cervical spine #2 D. cervical spine #7
C. cervical spine #2
Rationale:
In the maintenance phase of ALL, whole brain irradiation may be administered using the helmet technique where the inferior border is at C2.
An advantage of using the wedged pair technique is that beyond the point of beam intersection, there is:
A. rapid dose buildup
B. rapid dose falloff
C. isodose uniformity
D. heel-effect dose
B. rapid dose falloff
Rationale:
When using the wedged pair, the highest dose concentration is near the hinge, and then there is rapid dose falloff beyond the hinge.
A patient is receiving 3000 cGy total dose to a spine port using the SSD technique. The total given dose is 3636 cGy. What is the percent depth dose?
A. 100% B. 120% C. 82.5% D. 75%
C. 82.5%
Rationale:
Use the applied dose formula. Given dose = TD/%DD. 3636 = 3000/x; 3000/3636 = 0.825.
Monitoring setup reproducibility:
A. is usually accomplished by weekly port films
B. is always accomplished by using electronic portal imaging
C. can be accomplished by visual comparison of simulation films/DRRs and port films, or by electronic portal imaging
D. is not necessary in most patients
C. can be accomplished by visual comparison of simulation films/DRRs and port films, or by electronic portal imaging
Rationale:
Periodic comparison of treatment verification ports to DRRs or conventional simulation films is a means to monitor setup reproducibility.
The isodose chart shown is for an electron beam energy of approximately:
A. 5 MeV B. 7 MeV C. 9 MeV D. 12 MeV
C. 9 MeV
Rationale:
Using the rules for electron energy and isodose depths, you can apply the 80% rule: E/3 = 3 cm (depth of the 80% dose line).
What thickness of cerrobend would be equivalent to 5 cm of lead?
A. 4.2 B. 5 C. 6 D. 7.2
C. 6
Rationale:
Cerrobend is about 15% less dense than lead. To figure the equivalent attenuation of cerrobend to lead, use the factor 1.2. 1 cm lead = 1.2 cm cerrobend.