Module 6 Flashcards

1
Q

According to MQSA standards, the collimated field may not exceed the image receptor on any side except the _______ side.

nipple
right
left
chest wall

A

chest wall

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

The most common source of film fogging in mammographic darkrooms is _______.

use of out dated chemicals
radiation exposure
poor air circulation
poor lighting conditions

A

poor lighting conditions

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

A STEP test assesses:

automatic compression functions
film processing equipment and conditions
darkroom cleanliness only
automatic exposure control levels

A

film processing equipment and conditions

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

A film-screen contact test should be performed ________.

annually
semi annually
quarterly
monthly

A

semi annually

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

An outcomes audit must be initiated within _____ months after a facility becomes certified.

3
6
12
18

A

12

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Baseline QC standards for mammographic processors include measurement of all of the following except __________.

base-plus-fog density
mid-density
max density
density difference

A

max density

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

The crossover procedure for mammographic film processing is used to adjust the QC procedure to accommodate new __________.

processing temperatures
film batches
processing speeds
screen speeds

A

film batches

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

MQSA requires cassette screens be cleaned at least ___________.

daily
weekly
monthly
quarterly

A

weekly

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

The technical factors used when imaging a test phantom should be:

specified by the phantom supplier
the same as those used for the densest breast
the same as those used for a normal breast
any factors set by the tester, as long as they are documented

A

the same as those used for a normal breast

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

When scoring a phantom image, the technologist can see 3 full fibers and 1/2 of the fourth fiber. The score for fibers in this scenario is ___________.

2.5
3.0
3.5
4

A

3.5

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

The MQSA requires the facilities perform an image quality test using an FDA approved phantom at least ___________.

quarterly
monthly
weekly
daily

A

weekly

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

According to the MQSA, corrective action must be taken within 30 days of repeat analysis when a facility’s repeat rate exceeds _______ %.

2
3
4
5

A

5

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Hypo test solution is used for a __________ test.

darkroom fog
screen cleanliness
fixer retention
film-screen contact

A

fixer retention

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

When performing a darkroom fog test, the 2 measurements on the finished image should be taken:

one each inside and outside of the phantom image.
inside the phantom test object image.
outside the phantom test object image.
on an actual mammogram; no phantom is used.

A

outside the phantom test object image.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Maximum compression force on a mammography unit must not exceed ________ newton.

25
45
111
200

A

200

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

The evaluation of system resolution is sometimes referred to as focal spot performance.

True
False

A

True

17
Q

Film artifacts that appear parallel to each other when fed through a processor originate in the __________.

phantom image
processor
mammography unit
film base

A

processor

18
Q

MQSA regulations state that kVp must be accurate to within + ________% of the selected kVp at the lowest setting.

2
5
7
15

A

5

19
Q

The view boxes used for interpreting mammography images should produce at least _________ candela per sq meter luminance.

1000
1500
2000
3000

A

3000