Measurement Devices Flashcards

1
Q

Fowler’s Method

A

Measurement of anatomical dead space. Single breath N2 washout. Breathe in 100% O2 (single breath) then plot N2 concentration.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Absorption wavelengths of Pulse Oximeter

A

660nm (red) and 940nm (infrared) - then the ratio of absorption is converted via calibrated normograms

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Absorption characteristics of MetHb

A

Absorbs more IR than OxyHb or DeOxyHb therefore results in an R value approaching 1, causing a calculated SpO2 trending towards 85%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Isobestic Point of Pulse Oximetry

A

805nm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Absorption characteristics of COHb

A

Predominantly absorbs light at 660nm and does not absorb at 940nm therefore causes a falsely raised reading

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

CPEX: Anaerobic Threshold

A

Less than 11 ml/kg/min = poor outcomes

How well did you know this?
1
Not at all
2
3
4
5
Perfectly