Magnetism - Hardware Control Flashcards
High field scanners typically have field strengths greater than or equal to:
1.5 Tesla
Consistent room temperature of 65-75°F, humidity of 50-70%, and sufficient air flow in the MR scan room is recommended to:
Make patients less vulnerable to the RF heating effects
In a cross sectional view of the MR magnet design, which coils are closest to the patient?
Radiofrequency coils
The RF coils are closest to the patient, leading to the primary component causing RF absorption in the patient
The liquid cryogens are cooled in a superconducting magnet at a temperature of _________?
4° Kelvin
Which of the following is NOT a magnetic field associated with MR imaging?
Gravitational earth magnetic fields
The purpose of shimming the magnet is to:
make the B0 field as homogenous as possible
The array processor is responsible for:
Reconstructing the images using the Fourier transform
The array processor is responsible for reconstructing the collected MR images/data using the Fourier transform, and is located in the MR equipment room.
In order for energy to be efficiently transferred between a transmitter and receiver, the two must be at the same:
Frequency
In order for energy to be efficiently transferred between a transmitter and receiver, the two must be at the same frequency
The _____ is located within an equipment room.
RF power amplifier
The RF power amplifier is located within an equipment room
During a quench of the MR system, patients and employees should be evacuated from the scan room to avoid ______________.
All of the above
Field strength at magnet isocenter is measured in units of:
Tesla
Field strength at magnet isocenter is measured in units of Tesla, the SI derived unit of magnetic field measurement.
The unit of measurement of the magnetic field surrounding the periphery of the MR scanner is expressed as:
Gauss
The unit of measurement of the magnetic field surrounding the periphery of the MR scanner is expressed as Gauss.
The intensity limit for a static magnetic field in clinical use is limited by the Food and Drug Administration (FDA) to _____________?
4 Tesla
The intensity limit for a static magnetic field in clinical use is limited by the Food and Drug Administration (FDA) to 4.0 tesla (T) strength.
The patient image display with sequence parameter visualization is located in the _________.
Scan control room
The patient image display with sequence parameter visualization is located in the scan control room on the operator console.
The local RF coil used to image a human brain is typically located/kept in the __________.
Magnet room
The local RF coil used to image a human brain is typically located/kept in the magnet room.