9/22/22 Flashcards
- At 1 Tesela the frequency difference between that of fat and water is approximately
147 Hz
- How well a material attracts the imaginary lines of the magnetic field is defined as:
Magnetic Permeability
- _______ occurs after initial RF application, causing phase differences to appear between precessing spins, resulting in decay of transverse magnetization.
Dephasing
- The difference in chemical shift is approximately ___ parts per-million
3.5
- Aside from the center frequency, what other indicator in Image 82 signifies it was acquired at 1.5 Tesla?
The chemical shift between fat and water is 220 Hz
-The net magnetization vector (NMV), when aligned with the magnetic field direction, is aligned ________.
Along the longitudinal axis
- When thermal equilibrium is reached:
There are more hydrogen protons in the low energy state
- At 1.5 Tesla, the chemical shift between fat and water is:
220 Hz
Image 60 represents the ( w0 = y x B0):
Larmor equation
-Hydrogen protons in the anti-parallel state are referred to as:
Both A and C spin down and/or high energy spins
-Immediately following the application of the 90degree RF pulse, the precessing protons:
Both A and B begin to precess in phase and tip into the transverse plane
-Immediately following the application of the 90degree RF pulse, the transverse magnetization is:
Non-zero
-In one T2 relaxation time:
63% of the transverse magnetization has decayed
-In one T1 relaxation time:
63% of the longitudinal magnetization has recovered
______ magnets can be “shut off” the quickest in case of an emergency
Resistive
A magnetic vector possesses attributes of both _____ and _____
Magnitude, direction
Calculate the Larmor frequency for H at 0.5 Tesla.
21.28 Mhz
All the following are TRUE except:
The RF field oscillates at the Faraday frequency
What is defined as the excess number of hydrogen protons aligned with the static magnetic field direction (B0)?
Longitudinal magnetization
The energy used to form MRI images is _____ the patient’s tissues.
Emitted from
The extent to which a material or tissue becomes magnetized in an external magnetic field is called:
Magnetic susceptibility
If the RF energy transmitted is ___ Mhz, the slice through the patient would be in a magnetic field of .995T.
42.36
The precessional frequency of Hydrogen according to the Armor equation at 3T is:
127.71 Mhz
The protons that align themselves with the direction of the static magnetic field (B0) exist in a ____ energy state that those anti-parallel to the magnetic field.
Lower
The first documented human MR image was obtained on a 15 Mhz RF system, what field strength did it operate at?
0.352 Tesla
What would the scan time be in a 3D volume acquisition as follows: TR 32ms, TE 4 ms, Flip Angle 15degrees, FOV 18cm, 2mm slice thickness, 96 slices, 192x256 matrix, 1 NEX?
9 min 50 sec
3D scan time formula: TR x Phase Matrix x NEX x #slices
Calculate the voxel volume for the following pulse sequence parameters: TR 450, TE 20, FOV 24cm, 320x320 matrix, 3mm slice thickness.
1.69mm3
To calculate voxel volume, first determine pixel area. FOV/matrix xFOV/matrix = pixel area. Pixel area x slice thickness = voxel volume.240/320 x 240/320 = 0.5625 mm2. 0.5625 x 3 = 1.69mm3
Calculate the voxel volume for the following pulse sequence: TR 2000, TE 85, FOV 28cm, matrix 272 x 320, 5mm slice thickness, 3 NSA.
4.55 mm3
Calculate the pixel area for the following sequence: TR 450, TE 12, FOV 18cm, 224x256 matrix, 4 NEX.
0.56mm2
FOV/matrix x FOV/matrix = pixel area
______ is when only half the views of k-space are filled in the frequency axis.
Partial or fractional echo
A method of reducing MRI scan time by not filling as many lines of k-space in the phase direction, while maintaining spatial resolution, is known as:
Rectangular FOV
The overall signal-to-noise ratio will increase in 3D volume imaging when there is an increase in:
A and/or D increase in either FOV and/or increase in the number of slices or partitions