Lexi's Physics Qs Flashcards
When using a flow-directed, balloon-tipped, pulmonary artery catheter (Swan-Ganz)
- There is direct measurement of the left atrial pressure
- Once in place, the balloon is left inflated to prevent displacement
- Wedge pressure is approximately equal to the left ventricular end diastolic pressure
- The volume of the balloon is 5ml
- Pressure should be measured at the end expiration
FFTFT
The following methods have been used to assess depth of anaesthesia
- Isolated forearm technique
- Sensory Evoked Potentials
- Lower oesophageal sphincter contractility
- Train of Four Stimulation
- Transthoracic impedence measurement of cardiac output
TTFFF
The hazard of micro shock in hospital can be reduced by the use of
- Isolated (floating) power supply
- Saline-filled intracardiac catheters
- Battery powered appliances
- Multiple earth paths
- Large area diathermy plates
TFFFF
The following are true
- The application of logarithms will simplify all arithmetic manipulations
- A logarithmic function is a power function
- A graphical representation of y=ax+c is a straight line
- A clinical measurement that is used in monitoring is termed a parameter
- If the product of two variables is a constant then the relationship is hyperbolic
FTTTT
In a time cycled ventilator providing constant pressure generation during the inspiratory phase
- A source driving gas at four bar is required
- The tidal volume is largely unaffected by leaks around the trachea tube
- And exponential inspiratory flow waveform is produced
- The tidal volume is unaffected by changes in lung compliance
- The peak inspiratory pressure is an indication of airway resistance
FTTFF
Humidity may be measured by
- Mass spectrometry
- Chromatography
- Weighing
- Changes in length of biological tissue
- Detection of a dew point
TTFTT
Helium
- Has a similar viscosity to oxygen
- Is stored as a liquid in cylinders
- When inhaled causes voice changes
- Usefully decreases the work of breathing in bronchospasm
- Support combustion
TFTTF
Blood transfusion
- May result in hypokalaemia with rapid transfusion in hyperkalaemic patients
- May result in hypocalcaemia
- Should be warmed to 42°C
- May result in metabolic alkalosis
- PH of stored blood is 7.4
TTFTF
The systolic BP maybe overestimated by sphygnomanometry if
- The patient’s arm is very fat
- The cuff deflated too slowly
- Measurements are made at half minute intervals
- Too narrow a cuff is used
- The sphygmomanometer is at a higher level than the arm
TFFTF
TEC 6 vaporisers
- Requires an electrical supply
- Can be used for Sevoflurane
- Are accurate with flow rates for up to 10 L/m
- Cannot be filled when in use
- Are calibrated up to 18% of volatile agent
TFTFT
The laryngeal mask (LM)
- Is safe to be used in patients with latex sensitivity
- Is suitable for use in neonates
- Allows near-normal respiratory physiology with spontaneous ventilation
- Can be accidentally inserted into the oesophagus
- Can be used safely with a full stomach
TTTFF
Lasers are safe to use for an ENT procedure with
- Nitrogen
- Jet ventilation
- A standard Flexi-metallic tracheal tube
- A tracheal cuff filled with water
- Helium
TTFTT
With surgical diathermy
- Direct current is required
- Current frequency is typically more than 1 MHz
- Heat produced is proportional to current density
- Bipolar diathermy is more powerful than monopolar diathermy
- Bipolar diathermy is contraindicated in patinets with a cardiac peacemaker
FFTFF
Lung function tests
- An FEV1/FEV ratio of more than 65% is normal
- Carbon Dioxide is used to calculate the gas transfer factor
- FRC can be measured by helium dilution spirometry
- Closing capacity is always larger than closing volume
- Total lung capacity can be measured by plethysmography
FFTTT
With regards to monitoring of neuromuscular block
- The diaphragm and laryngeal muscles are more sensitive to relaxants than the abductor pollicis muscle
- The diaphragm and laryngeal muscles recover more quickly than the abductor pollicis muscle
- The nerve stimulator used to monitor neuromuscular block delivers a biphasic wave
- Double burst ‘fade’ is more readily detectable by visual means than train-of-four fade
- Depolarizing block is characterized by post-tetanic potentiation
FTFTF
-Recovery of neuromuscular function at the end of any procedure is best reflected at the adductor pollicis muscle where neuromuscular transmission is last restored