LDU Managers Module Flashcards
What is the flow line?
Way instruments should be processed around a single room LDU
Minimises risk for cross contamination
Dirty to clean
What is the instrument life cycle?
Acquisition
Cleaning
Disinfection
Inspection
Packaging
Sterilisation
Transport
Storage
Use
Transport
Daily WD tests?
Check spray arms and filter
Check door seal
Check chemical level
Check load carrier
Record cycle data
Loading patterns WDs
No overlapping
Don’t overload
Assemblies disassembled
Forceps open at hinge
Loading patterns WDs
No overlapping
Don’t overload
Assemblies disassembled
Forceps open at hinge
Cycles of WD?
Flush - remove gross debris
Wash - remove biological matter
Rinse - remove residue
Thermal disinfection - 90/95 for 1 minute
Drying - remove moisture
What to check at ispection
Check contamination
Check function of instrument
Check damage or corrosion
Process of manual cleaning?
SOP
Immersion or non immersion
PPE
- heavy duty gloves, apron, face covering
Chemicals
- <35 degrees, soft bristles brush
Conditions of steriliser?
134-137 degrees for 3 mins
2.05 - 2.35 bar
Daily steriliser checks
Check load carrier and chamber
Check door seal
Fill feed water reservoir
Drain used water
Record cycle info
Tests for type B steriliser?
Bowie dick / steam penetration
Air leakage
Air detector function tests
Record cycle data