What 2 guidance docs are there for decontamination relevant to us?
SHTM 01 - 01 Decon of medical devices in central Decon unit
SHTM 01 - 05 management, equipment and process of the Decon of dental instruments in local Decon unit
Why are GLS good?
They’re developed according to expert scientific, healthcare and engineering collaboration
Inform all relevant individuals involved with installation and usage on proper maintenance and ongoing usage
Inform those using machines on how to operate for consistent performance
Where might one find info on the process of decontamination?
SHTM 01 - 01, part A
Identifies and explains stages required
Weekly tests for WD?
Daily tests
Water hardness
Water conductivity
Cleaning efficacy test
Quarterly tests WD?
Weekly and daily tests
Check calibration
Check chemical dose
Check thermometric tests
Door and door interlock mechanism tests
What are the exposure time, temperature bands for WD?
100 mins at 70-75 degrees
10 mins at 80-85 degrees
1 minute at 90-95 degrees
Which SHTM is available for WD and which for steriliser?
WD is SHTM 01 - 01 - part D
SS is SHTM 01 - 01 - part C
What is the suitable time / temp for steam steriliser
134 - 137 degrees
2.25 bar pressure
3 minutes