LDU Flashcards
what are the requirements for manual cleaning
30-35 degrees
60ml of chemical in 12L sink
non-wired brush
stages of the washer disinfector
pre-wash
main wash
rinse
thermal disinfection
drying
disinfection temperature and hold time on the washer disinfector
90-95
1 minute
daily checks for the washer disinfector
spray arms spin freely
spray jets not blocked
no debris on filter
door seal condition
amount of chemical
record disinfection temperature of first cycle every day
loading requirements of the washer disinfector
no overlapping
hinges open
disassemble things
temperature of sterilisation
134-137 degrees celsius
pressure of sterilisation
2.05-2.35 bar
hold time of sterilisation
3 minutes
daily tests of the steriliser
door seal
chamber free from damage and debris
condition of load carrier
fill and drain feedwater reservoir each day
drain used water reservoir each day
steam penetration
what is in the sinner circle
energy
temperature
chemicals
time
what is the guidance document for decontamination in dentistry
SHTM 01-05
what would the GUID 5005 tell you
how to lay out decon room
if you wanted to buy a decontamination machine what guidance would you look at
NP 143
what do instruments need to be marked with to know they are safe for use in the UK
CE mark
what is in the ACT of the WD
temperature for each stage
time each stage active for
disinfection start and end with stage duration
volume of chemical dosed
cycle time overall
what is in the sterilisation ACT
pressure
temperature at start of sterilisation and at one minute intervals until stage ends
steam penetration test
air leakage test
air detector function test
where can you find information on daily and weekly testing of LDU
SHTM