interferential current Flashcards
what is the Tx frequency and clinical goal of premodulated (high)?
60-150hz
acute pain
what is the Tx frequency and clinical goal of premodulated (high/low)?
0-150hz
sub-acute pain
what is the Tx frequency and clinical goal of premodulated (low)?
0-15hz
chronic pain
what is the treatment area of ALL premodulated interferential current?
small area
what frequency range would be used for acute pain in both premodulated and true interferential current?
60-150hz
what is the electrode placement of all premodulated interferential current?
one electrode on either side of the pain
what is the intensity and target fiber of premodulated (high) interferential current?
patient comfort (sensory only) A-beta
what is the intensity and target fiber of premodulated (high/low) interferential current?
patient comfort (sensory through muscle twitch) A-beta (sensory) and A-Delta (motor)
what is the treatment area of ALL true interferential current?
large area
EX. low back pain
what is the frequency and clinical goal of true interferential (high) current?
60-150hz
acute pain
what is the frequency and clinical goal of true interefential (high/low) current?
0-120hz
sub-acute pain
what is the frequency and clinical goal of true interferential (low) current?
0-15hz
chronic pain
what is the electrode placement of all true interferential current?
crisscross/clover leaf around area of pain
what is the intensity and target fiber of true interferential (high)?
patient comfort (sensory only) A-Beta
what is the intensity and target fiber of true interferential (high/low) current?
patient comfort (sensory through muscle twitch) A-beta (sensory) and A-delta (motor)
what is intensity and target fiber of True interferential (low) current?
patient comfort (muscle twitch) A-delta
what are the indications for interferential current?
pain reduction
- acute and chronic
what are the contraindications for interferential current?
anterior cervical area areas of hemorrhage abdomen, pelvis, lumbar areas of pregnant women cancer confused and unreliable patients damaged skin epilepsy metal implants pacemakers and implantable cardioverter-defibrilators through the head and chest unknown etiology