How to fix Artifacts Flashcards

1
Q

truncation

A

increase number K space lines (phase matrix)

causes increase in scan time

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

flow

A

use pre-sat pulses

increase to SAR

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

chemical shift

also, where does it occur

A

occurs along the frequency axis
increase RBW
decrease in SNR and minimum TE results

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

out of phase artifact/aliasing

A

use no phase wrap, or increase FOV

also increases scan time

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

zipper

A

call engineer for repair

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

magnetic susceptibility

A

use spine echoes
remove metal if possible
use parallel imaging

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

motion

A

pre sat pulses
gating
increased scan time results

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

cross talk

A

larger slice gaps

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

moire

A

use spin echoes

make sure they are not touching bore

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

magic angle

A

change TE

change position of anatomy being scanned

How well did you know this?
1
Not at all
2
3
4
5
Perfectly