homework review 2 Flashcards

1
Q

According to ACR Mammography Accreditation standards, what is the minimum frequency for evaluating the compression thickness indicator by the technologist?

A

monthly

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

During ACR phantom scoring, what will result in a full point being awarded when viewing the fibers?

A

> = 8 mm long

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

During ACR phantom scoring, what will result in a full point being awarded when viewing the specks?

A

4-6 specks

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

During ACR phantom scoring, what will result in a full point being awarded when viewing the masses?

A

≤½ & >¾ border

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Mammographers must perform phantom image evaluation at least

A

weekly

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

The MQSA does not require facilities to have

A

ACR accreditation

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

The average glandular dose delivered during a single craniocaudal view of an FDA-accepted phantom simulating a standard breast shall not exceed ____ mGy per exposure.

A

3

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

The mAs recorded during phantom image testing of mammographic equipment should not vary by more than ____ from the initial control value.

A

±15%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

The maximum allowable dose per exposure for mammographic images is ____ rad.

A

0.3

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

The recommended maximum tolerance of kVp variance for mammographic x-ray generators is

A

5%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

A visual inspection of a CT unit should be evaluated at least _____ for ACR accreditation programs.

A

monthly

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

For ACR CT accreditation, the Water CT Number and Standard Deviation (Noise) should be evaluated by the CT technologist

A

daily

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

For a slice thickness of 5 mm or less, the slice thickness should not vary by more than ____ mm.

A

0.5

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Patient dose values for CT scans should remain within ____ of the manufacturer’s specifications.

A

±10%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Section thickness should not vary by more than ____ mm for sections greater than 5 mm.

A

±1.0

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

The CT number of water should not deviate by more than ____ CT numbers from 0.

A

3

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

The CTDIvol pass/fail criterion for an adult abdomen is ____ mGy.

A

30

18
Q

The laser light field and radiation field should coincide to within ____ mm.

A

2.0

19
Q

The minimum distance between two objects that allows them to be seen as separate and distinct would define what CT imaging parameter?

A

high contrast resolution

20
Q

The uniformity of the CT number of water should not vary by more than ____ HU from the center of the phantom to the periphery

A

5

21
Q

A change in the strength of the main magnetic field can be caused by

a. All are correct.
b. external ferromagnetic materials.
c. cryogen boiloff.
d. shim coil failure.

A

a. All are correct

22
Q

An acquisition with a 200-mm field of view and 256 x and y encodes yields a pixel size of ____ mm2.

A

0.78

23
Q

Luminance uniformity of soft-copy displays used in MRI systems should be within ____ of the maximum brightness measured in the center of the screen.

A

30%

24
Q

Nonuniformity in the magnetic field can be caused by

a. All are correct.
b. the presence of ferromagnetic material within the bore of the magnet.
c. uniformity in the RF coil.
d. a homogenous main magnetic field.

A

b. the presence of ferromagnetic material within the bore of the magnet.

25
Q

The ACR slice position deviation in MR scanners should not exceed ____ mm.

A

5

26
Q

The ability of an MR scanner to resolve small objects is known as

A

high contrast resolution

27
Q

The ability of the MR system to differentiate between adjacent tissues having minimal differences in signal intensities is known as

A

low contrast resolution

28
Q

The amount of geometric distortion in the image is the

A

spatial linearity

29
Q

What is the maximum luminance for soft-copy display devices as suggested by the ACR?

A

greater than 90 nit

30
Q

What is the resonant frequency in a unit with a 1.0-T magnet and a gyromagnetic ratio of 42.57 MHz/T?

A

42.57 MHz

31
Q

A chemical fog artifact

a. appears as streaks.
b. results in lower contrast.
c. results in lower optical density.
d. is an exposure artifact.

A

b. results in lower contrast.

32
Q

Data for determining repeat rates should include at least ____ patients to obtain a statistical sample large enough for valid results.

A

250

33
Q

If residual thiosulfate is not properly removed during washing, the radiograph will

A

turn brown

34
Q

Patient motion during exposure results in _____ artifact.

a. a handling
b. no
c. a processing
d. an exposure

A

d. an exposure

35
Q

Pi lines are so named because they

a. occur at regular intervals.
b. appear as a section of pie.
c. None are correct.
d. occur on the finished radiograph.

A

a. occur at regular intervals

36
Q

Wet pressure sensitization would be classified as _____ artifact.

a. an exposure
b. no
c. a processing
d. a handling

A

c. a processing

37
Q

Which of the following artifacts can occur in both CR and film/screen radiography?

a. Moiré pattern
b. Water spots
c. Static electricity
d. Wet-pressure sensitization

A

a. Moiré pattern

38
Q

Which of the following can occur as a result of a double set of grid lines?

a. Grid misalignment
b. Film fog
c. Static artifacts
d. Moiré effect

A

d. Moiré effect

39
Q

Which of the following will run in the same direction as that of film travel?

a. Guide shoe marks
b. All are correct
c. Pi lines
d. Hesitation lines

A

a. Guide shoe marks

40
Q

_______ can appear as dark bands at the interfaces of structures that differ widely in brightness level such as barium examinations or metal prosthesis.

A

Halo artifacts