Homework Flashcards
After a patient performed an FVC, the therapist performed a validity
check, and noticed that the back extrapolated volume was 175 mLs.
What should the therapist do?
Invalidate the test
At the start of a PFT the therapist turns on the PFT machine, but the
machine will not actually turn on. What actions should the therapist
take?
The device lacks electrical power.
Confirm device:
Is plugged in
Wire connections are good
Check/replace batteries.
Then the portable spirometer turns on but does not complete or fails
the power-on self-test. What actions should the therapist take?
failure of boot/start-up program or central processing unit
failure.
Record error message.
Turn device off, wait 20 seconds, turn device on.
Replace device
During the 3 L volume calibration the portable spirometer reads 3.12
L? What actions should the therapist take?
(3090 L is the acceptable high for calibration)
Check temp/press of altitude input
Check/fix flow sensor assembly
Check flow sensor for obstructions
At the start of the PFT the therapist notices that volume accumulated
before the exhalation. What should the therapist do?
Have patient hold the sensor steady at the beginning of the test, or set
it on the table
What should the therapist do if the flow measures appear to be
reversed in the PFT results?
The flow sensor for the inlet and outlet pressure tubing is
reversed.
Check/correct and confirm proper tubing placement
The spirometer did not sense the beginning of exhalation, what should the therapist do?
Check the sensor for damage and the sensor pressure tubing assembly and connections. Confirm correct tubing connections and/or replace
sensor.
The therapist suspects a low volume reading on a PFT what should be
done?
Recalibrate the PFT machine
Check the temperature
Pressure, or altitude inputs
Check the flow sensor for damage, reassemble/replace, check for leaks
in the tubing
Use nose clips
Ensure lip seal
if the therapist suspects high or low percent normal computations for
a PFT what should be done?
Check patient data entry (age, height, gender)
Reenter correct patient data
A patient has a PFT. The first FVC is 4.32 L. The second FVC is
4.6 L. What should the therapist do?
The therapist should do up to 8 tests until the patient tires or the 2
largest values are between 0.150 L
What does a peak flow device measure?
Flow (obstruction)
What type of patient most commonly utilizes a peak flow device?
Asthma patients
How are normal values established for peak flow devices?
The patient uses the peak flow meter 1 to 2 times/day.
Establishes their own normal
What is the green zone?
80% to 100 % of patient’s normal.
Patient is doing well
What is the yellow zone?
50 to 80% of patient’s normal.
May need to increase medications, call DR.
What is the red zone?
Less than 50% of patient’s normal.
Go to the hospital!
Define forced vital capacity (FVC).
Volume of air exhaled after a maximum inhalation.
What does the FVC measure?
Capacity or volume.
What does a low forced vital capacity mean?
Patient is restricted
Explain why the FVC is important when determining obstruction
If the FVC is low the patient may not look obstructed.
If the FVC is high the patient may look obstructed, but is not
obstructed. The FVC and the FEV1 need to be compared
define functional residual capacity (FRC
The volume of gas remaining in the lungs at the end of a resting
exhalation.
Define residual volume (RV
The volume of gas remaining after at the end of a resting
exhalation
Define forced expiratory volume (FEV1):
Volume of air exhaled in the first second of exhalation
What does the FEV1 measure?
Flow L/sec
What does a low FEV1 mean?
large amount of Obstruction. Patients cannot move air out of the
lungs, which makes it hard to get air into the lungs.
List possible ways to fix missed triggering
increase sensitivity
Increase PEEP if there is auto-PEEP
List 3 possible ways to fix air trapping
Increase PEEP
Decrease respiratory rate
Decrease inspiratory time
List possible ways to fix a rise or slope problem
- Decrease inspiratory time
- Increase the rise time
List possible ways to fix a prolonged inspiratory phase problem:
- Decrease the inspiratory time
- Change the cycle time
List possible ways to fix “Beaking”
- Decrease PEEP
- Decrease PIP
- Decrease tidal volume
A therapist enters a patient room, and the ventilator is emitting a low
volume and a low-pressure alarm. What actions should the therapist
take?
Check ETT placement
Check for disconnections and/or loose connections in ventilator tubing
A therapist enters a patient room, and the ventilator is emitting a low
volume and a high-pressure alarm. What actions should the therapist
take?
Check to see if the patient needs suctioning
Water in the ventilator tubing
Check for a decrease in patient compliance (change to a pressure
mode)
A therapist enters a patient room, the ventilator is emitting a low
volume and a low minute volume alarm on a spontaneous mode.
What actions should the therapist take?
Assess the patient for distress
Check minute ventilation and respiratory rate
Check ETT placement
Check for disconnections and/or loose connections in ventilator tubing
If patient is distressed, change to a more supportive mode
A therapist enters a patient room, the ventilator is emitting a high-
respiratory rate and a high minute ventilation alarm on a spontaneous
mode. What actions should the therapist take?
Assess the patient for distress
Check minute ventilation and respiratory rate
If patient is distressed, change to a more supportive mode.