*Healthcare Technology Problem Solving – 25% Flashcards

1
Q

What should a tech do when a ventilator fails and a nurse wants it repaired immediately?

A

Follow hospital policy

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Infant incubator isn’t working in skin mode when the infant is wrapped in a blanket. What is the most likely issue?

A

poor probe contact with skin

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

A biomed is going to perform a PM on a central station which will cause critical alarms. What should the biomed do first?

A

Notify appropriate unit coordinator or nurse

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Primary advantage of fluid column ECG monitoring electrodes is the ________.

A

compensation for patient movement

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

ECG cart data transmission to archiving system could be incomplete due to defective network ________ or improper network ________.

A

hardware, configuration

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Purple screen on a video monitor could be caused by.

A

Green component cable incorrectly attached

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Balloon pump alarms: “failure to inflate” or “low gas” can be caused by ________ or ________.

A

rupture or leak

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

OR surgery table is plugged in and on. Hand pendant has lights, but table all or most table movements don’t move. What is a common cause?

A

stabilizing feet not locked

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Movement, ambient light, cold extremities, NIBP pump on same extremity, diaphoresis, patient clutching hand or making fist, or fingernail polish can cause errors in what measurement?

A

SPO2

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

IV pump error caused by improperly blocked or crimped line near IV site.

A

Occlusion/distal alarm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

A surgery tower has video image, but slave doesn’t. Slave is hooked up through a printer. Cause could be?

A

Likely that wrong input is selected or printer is in menu mode.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

What could cause an ultrasound probe to heat during an exam?

A

lack of coupling media

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

If an NIBP monitor continuously pumps, what is the likely cause?

A

Leaky cuff or hose

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

A biomed is called to the OR for an anesthesia machine. The doctor wants it repaired immediately. What should the biomed do?

A

Follow hospital policy.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

If there is a low gas alarm on an intra-aortic balloon pump, what should you check first?

A

check catheter for blood

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

For invasive blood pressure measurement. If the tip of the catheter is lower than the transducer, then the blood pressure will read?

17
Q

Half of telemetry monitoring is lost at a central station. Most likely cause is?

A

Access point or antenna issue.

18
Q

Nurse states that surgical laser won’t turn on after housekeeping was in room but it worked in the last case. What is an easily resolved possible cause?

A

Emergency stop depressed while cleaning.

19
Q

If a water bath is not heating, but has power, what is the likely issue?

A

water level low

20
Q

PCA pump error of “dose not requested” is displayed. What could be the issue?

A

faulty controller

21
Q

What could defibrillator paddle pitting cause?

A

patient burns

22
Q

Simplest cause of no video capture on Endo computer from C-arm.

A

video input selection

23
Q

Lead I will display, but not Lead II or Lead III. What is the likely issue?

A

LL (Red) connection/lead bad