Essentials of perioperative nursing textbook questions Flashcards
AORN’s Patient Focused model includes which of the following domains?
A. Patient safety, physiologic response, behavioral responses, the health system
B. Patient teaching, patient safety, behavioral responses, discharge planning
C. Patient safety, patient assessment, discharge planning, the health system
D. Patient assessment, plan of care, discharge planning, the healthy system
A. Patient safety, physiologic response, behavioral responses, the health system
The perioperative nurses begins the patient’s care plan:
A. Prior to the procedure, based on information about the patient from the surgeon and other healthcare providers
B. In the holding area based on interview and assessment data
C. Prior to the procedure based on knowledge of the planned procedure, typical related nursing diagnoses, and resources required
D. When the patient enters the operating room and all attention is focused on supporting the patient
C. Prior to the procedure based on knowledge of the planned procedure, typical related nursing diagnoses, and resources required
Which of the following is a true statement about the RNFA?
A. An RNFA is an advanced practice perioperative nurse, regardless of his or her academic level of preparation
B. the RNFA position is an expanded role in perioperative nursing
C. the RNFA practices under the license of the physician
D. The RNFA must have an advanced degree in nursing
B. the RNFA position is an expanded role in perioperative nursing
The dose and timing of prophylactic antibiotic administration is determined by:
A. Protocol, which states the dosage and specifies that it be given 1 hour before surgery
B. The attending surgeon
C. the anesthesia provider
D. the optimal level of antibiotic in the patient’s serum and tissue at the time of surgery
D. the optimal level of antibiotic in the patient’s serum and tissue at the time of surgery
Three protocols that TJC implements in an effort to prevent wrong site surgery are:
A. Site marking, time out, and better documentation
B. time out, better documentation and consent form
C. Pre procedure verification, site marking, and time out
D. Site marking, consent form and time out
C. Pre procedure verification, site marking, and time out
Disinfection differs from sterilization in what way?
A. High level disinfection and sterilization are the same
B. Sterilization kills all spores; disinfection does not
C. High level disinfection kills vegetative bacteria; sterilization does not
D. Disinfection and sterilization both kill microorganisms, but only sterilization may be used for all medical devices
B. Sterilization kills all spores; disinfection does not
Critical items:
A. contact sterile tissue and the vascular system
B. Must undergo high level disinfection
C. Do not penetrate mucous membranes
D. include thermometers, crutches and BP cuffs
A. contact sterile tissue and the vascular system
For an item to be sterilized
A. It must be sequentially wrapped and placed upright in the sterilizer
B. Steam must penetrate the packaging completely and the intended parameters of moisture, temperature, and time must be met
C. It must be subjected to unsaturated steam for the appropriate amount of time and under the correct amount of pressure
D. The temperature must remain at 250 degrees Fahrenheit at atmospheric pressure for the proper amount of time
B. Steam must penetrate the packaging completely and the intended parameters of moisture, temperature, and time must be met
Which statement is true about steam sterilization? SELECT ALL THAT APPLY
A. Moist heat under pressure is economical
B. Steam sterilization denatures and coagulates protein
C. steam sterilization occurs at atmospheric pressure
D. Steam sterilization is a function of time and temperature
A. Moist heat under pressure is economical
B. Steam sterilization denatures and coagulates protein
C. steam sterilization occurs at atmospheric pressure
Which statement is true about autoclaves? SELECT ALL THAT APPLY
A. Air trapped in the chamber will interfere with sterilization
B. Sterilization begins only when the correct temperature has been reached
C. Cups and basins must be placed so that water will collect in them and form steam
D. Wrapped packages should be removed from the sterilizer immediately and allowed to cool for 2 hours
A. Air trapped in the chamber will interfere with sterilization
B. Sterilization begins only when the correct temperature has been reached
How might trapping of air in a steam autoclave interfere with sterilization?
A. the presence of air forces the temperature above the target range
B. Air in the chamber prevents the sterilizer from reaching the target pressure
C. Air interferes with the humidity level in the chamber
D. Air pockets prevent steam from contacting all surfaces of an item, compromising the sterilization process
C. Air interferes with the humidity level in the chamber
Which statement is true about steam sterilizers? SELECT ALL THAT APPLY
A. A Bowie-Dick test is done once each day in all gravity-displacement sterilizers
B. A prevacuum sterilizer is more efficient and preferred over a gravity-displacement sterilizer
C. Pre vacuum sterilizers are used for liquids
D. Pre vacuum sterilizers are less affected by incorrect arrangement of objects in the chamber
B. A prevacuum sterilizer is more efficient and preferred over a gravity-displacement sterilizer
D. Pre vacuum sterilizers are less affected by incorrect arrangement of objects in the chamber
Which statement is true about IUSS? SELECT ALL THAT APPLY
A. IUSS is not intended for sterilizing sets of instruments
B. Items sterilized with IUSS cannot be stored or held for use on a future case
C. Drying time is the same in IUSS as with regular sterilization
D. In an emergency, items can be rinsed in the scrub sink and sterilized with IUSS
A. IUSS is not intended for sterilizing sets of instruments
B. Items sterilized with IUSS cannot be stored or held for use on a future case
Which statement is true about IUSS? SELECT ALL THAT APPLY
A. IUSS is perfect for implants that arrive immediately before a procedure
B. IUSS is appropriate only in urgent clinical situations
C. IUSS is appropriate when there is only one specialty set of instruments
D. IUSS can only be done in a pre vacuum sterilizer
B. IUSS is appropriate only in urgent clinical situations
Which statement is true about IUSS? SELECT ALL THAT APPLY
A. Only devices with manufacturer’s instructions for IUSS should be sterilized in this manner
B. there is a risk of contamination during transport following IUSS
C. Items for IUSS cannot be wrapped
D. A rigid container should be used for IUSS
A. Only devices with manufacturer’s instructions for IUSS should be sterilized in this manner
B. there is a risk of contamination during transport following IUSS
What are some advantages of H2O2 sterilization? SELECT ALL THAT APPLY
A. Items to be sterilized can be wrapped in any material
B. There is a single cycle for all items
C. the process is an efficient alternative to EtO sterilization
D. cycle times are preset and the sterilizer is easy to operate
C. the process is an efficient alternative to EtO sterilization
D. cycle times are preset and the sterilizer is easy to operate
The spore used to monitor H2O2 sterilization is: A. Bacillus atrophaeus B. Geobacillis atrophaeus C. Geobacillus stearothermophilus D. Bacillus subtilis
C. Geobacillus stearothermophilus
Which packaging materials can be used for H2O2 sterilization? SELECT ALL THAT APPLY A. Muslin wrappers B. Tyvek pouches C. Polypropylene wraps D. Mylar pouches
B. Tyvek pouches
C. Polypropylene wraps
D. Mylar pouches
Which statement is true about EtO sterilization? SELECT ALL THAT APPLY
A. EtO is a toxic gas that must be managed according to strict regulations
B. EtO is appropriate for items that cannot tolerate the temperature and moisture of steam sterilization
C. The essential parameters of EtO sterilization are gas concentration, temperature, humidity and exposure time
D. Aeration must be done for all loads sterilized using EtO
A. EtO is a toxic gas that must be managed according to strict regulations
B. EtO is appropriate for items that cannot tolerate the temperature and moisture of steam sterilization
C. The essential parameters of EtO sterilization are gas concentration, temperature, humidity and exposure time
D. Aeration must be done for all loads sterilized using EtO
What are some advantages of EtO sterilization? SELECT ALL THAT APPLY
A. The process is rapid
B. the process is safe for items that cannot tolerate high heat and humidity
C. EtO is noncorrosive
D. EtO effectively penetrates large bundles and permeates all porous items
B. the process is safe for items that cannot tolerate high heat and humidity
C. EtO is noncorrosive
D. EtO effectively penetrates large bundles and permeates all porous items
Which statement is true about liquid chemical processing? SELECT ALL THAT APPLY
A. It is a low temperature, nonterminal process
B. Items must be used immediately following processing; they cannot be packaged or stored for later use
C. Only one flexible endoscope can be processed at a time
D. Peracetic acid disrupts protein bonds and cell systems, causing immediate cell death
A. It is a low temperature, nonterminal process
B. Items must be used immediately following processing; they cannot be packaged or stored for later use
C. Only one flexible endoscope can be processed at a time
D. Peracetic acid disrupts protein bonds and cell systems, causing immediate cell death