Equipment Maintenance Flashcards
1
Q
How often should machine be serviced
A
At least annually
2
Q
How do you test cassette for light leakage
A
Load & leave in light for 1hr
Develop film - if blackened at edges, cassette needs replaced
3
Q
How often should cassettes be cleaned
Process of cleaning
A
At least monthly
Damp swab w/mild soap or specialised cleaner Wipe from centre to outside Stand vertically to air dry Label w/date and initials Reload new film
4
Q
Care & storage of films
A
Mark delivery date on box Store in dark room Store films in light-proof container upright - only open in safe-light conditions - away from x-ray area Must be used within date Avoid handling film
5
Q
Automatic processed warm-up length and what is done during this period
A
~10-20mins
Old, clean processed film passed through
- assess functioning
- remove dried chemical residue from rollers
6
Q
Films processed prior to reaching operating temp in automatic processor…
A
Underdeveloped