Equipment Maintenance Flashcards

1
Q

How often should machine be serviced

A

At least annually

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

How do you test cassette for light leakage

A

Load & leave in light for 1hr

Develop film - if blackened at edges, cassette needs replaced

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

How often should cassettes be cleaned

Process of cleaning

A

At least monthly

Damp swab w/mild soap or specialised cleaner
Wipe from centre to outside 
Stand vertically to air dry
Label w/date and initials
Reload new film
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Care & storage of films

A
Mark delivery date on box
Store in dark room
Store films in light-proof container upright 
- only open in safe-light conditions
- away from x-ray area
Must be used within date
Avoid handling film
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Automatic processed warm-up length and what is done during this period

A

~10-20mins

Old, clean processed film passed through

  • assess functioning
  • remove dried chemical residue from rollers
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Films processed prior to reaching operating temp in automatic processor…

A

Underdeveloped

How well did you know this?
1
Not at all
2
3
4
5
Perfectly