E-0016 - Enterolert Flashcards
What is the code for Enterolert?
SM 9230 D
What is the fist thing you do?
Make sure I have proper PPE on
- Lab coat, goggles, and gloves
What do you do after putting on PPE?
Turn on the Quanti-tray sealer machine
- Warm up for 10-15 minutes
- Ready when green light is on
What do you do after turning on the Quanti-tray sealer machine?
Prepare the logbook
Preparing the logbook: what information do you include?
What do you do after preparing the logbook?
Sterilize the work bench area, equipment, and gloves with 70% IPA
Describe the setup for an Enterolert.
- In rack:
- 2 QC bottles at the top: EF and EC
- Samples along the bottom, left to right, 5 in a row - Paper towels
- Lay media packs on them, plastic side up - Take out QT packets and label them with the sample description on the back
What do you do after sterilizing work area/equipment/gloves?
Check sample volume (should be 100 mL)
- If exceeds: pipette off excess
- If less than 95 mL: tell PM immediately
- If overfilled: transfer to sterile container to homogenize and transfer back to sample bottle
What must you do before pipetting off excess sample volume?
Shake the sample 25 times
What do you do after checking the sample volumes?
Get Enterolert media
What do you do after getting Enterolert media?
Spray the front of packet (plastic side) with 70% IPA
What do you do after sterilizing the media packets with 70% IPA?
Add the media to the samples and QC
- Gently swirl and invert to dissolve
What do you do after adding the media to the samples?
Pour the samples into the corresponding packets
- Tap the wells to release the air bubbles
What do you do after pouring the samples into the packets?
Run the packets through the sealer
- Place packet into rubber insert
- Run tray through the sealer
- Check for leaks after
- Sterilize the rubber insert after and between packets
What do you do after sealing the trays?
Incubate the samples and QCs