What quality does precision assessment measure?
reproducibility
How much radiation does DXA expose patients to relative to background radiation
less than 1 days worth of background radiation
Which skeletal site offers the highest precision and most rapid response to treatment?
lumbar spine
When should cross-calibration be performed?
when changing DXA hardware
To cross-calibrate DXA equipment, 30 patients representing the facility’s population should be scanned once on the old system and twice on the new system within how many days?
60 days
Serial scanning of a patient’s bone mineral density values can be used to monitor what?
response to therapy
Quality control includes any process or system of measures with what goal?
assessing quality
What is the main purpose of Stewart rules as they apply to bone densitometry?
Evaluating scanner precision
Spine phantoms used in quality control testing for DXA scanners are called?
anthropomorphic phantoms