DR QC Flashcards
Developed a pettern for evaluating display resolution
Society of Motion Picture and Television ENgineers (SMPTE)
SMPTE
Most common is the observation of (1) and (2) luminance patches to identify (3) in luminance
- 5%
- 95%
- gross deviation
recommended the SMPTE pattern for QC in telerdiology
American College of Radiology (ACR)
created the DICOM stndard
ACR and National Electrical Manufacturers Association (NEMA)
For consistent grayscale appearance in DICOM-compatible displapys
Gray Scale Display FUnction
seved as an acceptance testing standard for digital displat devices
DIN 2001
Provides evaluation instructions for system performance in line with compliane standards
Flat Panel Displat Measurement (FPDM)
Released FPDM
Video Electronics Standard Association (VESA), 1998
Evaluates a digital display device comprehensively to ensure acceptable clinical performance
AAPM TG 18
For luminance response of monitors and luminnace uniformity measurements
Photometer
Photometer types
Near range photometers
Telescopic photometers
Used in close proximit to monitors
Near-range photometers
Measure luminance from a distance of 1m
Telescopic photometer
Photometers required range: (1) with better than (2) accuracy
and at least (3) precision
- 0.05-1000 cf/m^2
- 5%
- 0.01
Evaluates monitor display reflection and ambient light conditions
Illuminance meter