CT Tutor Test 1 Wrong Answers Flashcards
Cone beam artifacts are related to the advances found in multi-row scanners because:
The traditional filtered back projection method of reconstruction cannot handle image formation with a beam collimation greater than 20mm
The system that consists of a secure network for transmission of patient information, a CT scanner, multiple workstations for interpreting and reviewing images and storage archived is called:
PACs
SFOV (or scan field of view) controls the diameter of the data acquired and is determined by
The detector array
The positioning system that assumes the patient is in the supine position is called:
Patient coordinate system
The technology that allowed the smaller, high-frequency generator to be located inside the CT gantry:
Slip ring technology
The main purpose of filtered backprojection is
To create artifact free images
What is the normal range for adult blood urea nitrogen (BUN) levels?
5-20 mg/dL
As the section (slice) width is increased the collimation is decreased.
True
False
True
Collimation determines slice width, the greater the collimation the smaller the slice width.
Lymphoma
Which of the following symptoms would be exhibited in a patient experiencing shock?
Rapid, shallow breath
Cyanosis
Tachycardia
All of the above
All of the above
Of the following x-ray interactions between the x-ray beam and the patient which results in the most patient dose?
Photoelectric effect
Bremsstrahlung
Compton scatter
Characteristic
Photoelectric effect
The photoelectric interaction results in the complete absorption of the energy from the incident x-ray photon within the patient’s body.
The type of image reconstruction used in the first-generation prototype CT scanner was _____________.
Convolution method
Fourier transform
Back-projection
Iterative technique
Iterative technique
With cupping artifacts, CT numbers are higher at the edges of the image and lower in the center.
True
False
true
There is less beam hardening at the edges of the image hence higher CT numbers and the radiation path are longer to the center of the image which causes a low number in the center.
Parotid gland
Submandibular gland
Sublingual gland
Salivary gland
Submandibular gland
If the filter at the aperture of the x-ray tube removed even more low energy x-ray photons which of the following would result?
An increase in partial volume artifacts.
A decrease in partial volume artifacts.
An increase in beam hardening artifacts.
A decrease in beam hardening artifacts.
A decrease in beam hardening artifacts.
Removing more of the low energy x-ray photons results in higher energy in the incident x-ray beam. This will decrease the amount of beam hardening artifacts due to the removal of the low energy x-ray photons.
When administering intravenous iodinated contrast media what type of infection control technique is required?
Surgical asepsis
Enteric precautions
Medical asepsis
Droplet isolation
Surgical asepsis
The average density of a solitary pulmonary nodule indicates it is benign when it falls within the following ranges…
15-30 Hounsfield units
50-80 Hounsfield units
125-145 Hounsfield units
175-200 Hounsfield units
175-200 Hounsfield units
Researchers feel pulmonary nodules appearing on CT examinations of the lungs can be considered benign if their average density is more than +164 Hounsfield units. The closer to +200 Hounsfield units the more certain the benign diagnosis is accurate.
Internal jugular vein
Internal carotid artery
External carotid artery
Common carotid artery
Common carotid artery
When completing a CT examination for the post-myelogram of the lumbar spine in an adult patient, which of the following would be an acceptable range for contrast volume for the intrathecal injection?
1-3 mL
5-9 mL
12-14 mL
18-22 mL
12-14 mL
The injection rate should not exceed approximately 17 mL in an adult patient, keeping the total dose of iodine below 3 grams. The most common range of injection during post-myelographic CT is 12-14 mL in an adult patient.