Collimation And The Beam Flashcards
1
Q
Beam restrictions
A
- device that attaches to the opening in the x-ray tube housing to regulate the size and the shape of an x-ray beam
- designed to control and minimize scatter by limiting the x-ray field size to only the anatomical structures of interest
2
Q
Purpose of beam restriction
A
- limiting patient exposure
- reducing scatter produced within patient
- increasing contrast on image
3
Q
True or false: beam restriction and collimation are synonymous
A
TRUE
4
Q
Off-focus radiation
A
- ideally all filament electrons are striking the anode at the focal spot
- but some stray electrons can stray, and interact on other areas of the anode
- this creates x-rays called ‘off focus radiation’
- this type of radiation if not ideal, increases image blur, which is not ideal for image quality
- ‘fuzzy edge’ of image: penumbra
- “zone of unsharpness”
- like blurry edge of a shadow
- the variable aperture collimator is designed to reduce off focus radiation
5
Q
Penumbra
A
- penumbra and off-focus radiation are reduced as the beam restrictor distance is increased from the tube port
- as the distance of the beam restrictor increases from the port the more the penumbra is reduced
- cylinder is more effective at reducing penumbra because it restricts the beam for a greater distance
6
Q
Beam restrictors
A
Three types
- aperture diaphragm
- cones/cylinders
- variable aperture collimator
7
Q
Aperture diaphragm
A
- simplest
- flat piece of lead or lead lined material that fits onto the x-ray tube head
- opening is usually just less than the size of the IR used
- field size controlled by: size of the opening, shape of the opening (circular, square, or rectangular) and SID
- advantages: inexpensive, easy to use
- disadvantages: need multiple diaphragms to achieve different field sizes, because it is so close to the focal spot, larger area of unsharpness around image (penumbra increased)
8
Q
Cones and cylinders
A
- modified aperture diaphragm
- extension cone or cylinder attached to a aperture diaphragm
- field size determined by the size and position of the distal end
- usually circular
- used only in special situations
- cone has a flared extension: upper diameter smaller than lower diameter
- cylinders have no flaring, same diameter from top to bottom
- advantages: inexpensive, easy to use, helps decrease penumbra around image (cylinder slightly more effective)
- disadvantages: fixed field size, shape of field is round, alignment is critical between x-ray source, cone, and IR, cones: if angle of flange is greater than angle of divergence of primary beam, not effective at restricting primary beam
9
Q
Variable aperture collimator
A
- best and most commonly used beam restrictor
- all modern general x-ray units utilize this system
- allows the most variation and flexibility in field size adjustments
- two sets of shutters, entrance shutters (limits the x-ray beam like an aperture diaphragm), purpose: to control off focus radiation
- equipped with a light source and a mirror
- light is intended to be an accurate representation of where the primary x-ray beam will be projected during procedure
- misalignment of mirror can cause anatomical structures to be cut off of the image
10
Q
What is an alignment test is a tech is concerned about the alignment of the light field and radiation field?
A
9 penny test
11
Q
Positive beam limitation (PBL)
A
- ‘automatic collimation’
- when an IR is placed in the bucky tray, automatic collimators mechanically adjust the size and shape of the radiation field to match that of the IR
- a way of protecting the patients from overexposure, but does not replace manual collimation by tech
- purpose is to ensure the field size does not exceed that of the IR
- can be built into APR (automatic programmed radiography), so when an exam is selected, the field size will automatically change
12
Q
Compton scatter interaction
A
- scatter is produced during a compton interaction
- scattered photons are not part of the useful beam
- scatter decreases the quality of the image by adding density over the entire image that isn’t related to the anatomy (losing contrast and appearing grey and dull)
- image fog
- no useful information
13
Q
Methods of reduction
A
Three primary factors that can influence the intensity of the scatter reaching the image
- kVp
- field size
- patient thickness
14
Q
Beam restriction
A
- as the x-ray beam is restricted through collimation, the amount of tissue being irradiated is decreased, results in: improved image quality, reduced patient dose
- as the field size is decreased, less scatter radiation is produced within the patient
- less patient anatomy is being irradiated, so fewer comptons interactions are occurring
- less scatter reaches the IR
- “more matter, more scatter”
15
Q
Field size and density
A
- when the field size is decreased, what is happening to the density on the image? Decreasing because:
- fewer photons reaching IR
- scatter, although not useful, contributes to the density of the image
- when collimating, some of the photons have been removed due to the field size: remember we don’t ‘squeeze’ the beam when it is collimated