Chapter 6 Quiz Flashcards

1
Q

The accuracy of the dose calibrator should be determined _______.

annually
quarterly
semiannually
daily

A

annually

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

The percent FWHM measures the _____.

-spatial resolution of the scintillation detector
-energy resolution of the scintillation detector
-linearity of the scintillation detector
-sensitivity of the scintillation detector

A

energy resolution of the scintillation detector

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Which of the following PET scanner calibration and quality control tests matches the activity measured in the dose calibrator to the radioactivity measured in the PET scanner?

-normalization
-blank scan
-absolute activity
-energy window calibration

A

absolute activity

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

How frequently should the blank scan be performed?

annually
quarterly
daily
semiannually

A

daily

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

The CT number calibration is checked daily by using a _______.

rob source
water phantom
flood source
blank scan

A

water phantom

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Which PET scanner calibration brings all detectors to provide the same sensitivity?

blank scan
energy window
normalization
coincidence timing

A

normalization

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

A PET/CT scan in the coronal plane demonstrates reduced intensity of FDG uptake in a uniform band above the liver. The most likely cause of this lack of uptake is ______.

-oral contrast in the gastrointestinal tract
-abscess
-breathing
-attenuating object

A

breathing

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Daily visual review of PET scanner performance is done by reviewing ______.

-axial slices of a uniform source
-sinograms
-sagittal/coronal plane review
-machine-generated written logs and reports

A

sinograms

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

The mathematical reproducibility test of a well counter is the _____.

-chi-square test
-energy auto calibration test
-linearity test
-geometry test

A

chi-square test

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Performance testing and licensing of CT scanners is required and conducted by _______.

-joint commission
-state/local agencies
-nuclear regulatory commission (NRC)
-national institute of standards and technology (NIST)

A

state/local agencies

How well did you know this?
1
Not at all
2
3
4
5
Perfectly