Chapter 18, 19 and 20 Study Guide Flashcards
Pulsed Doppler
range resolution
sample volume
limited maximum velocity-Nyquist
aliasing
Continuous Wave Doppler
range ambiguity
region of overlap
unlimited maximum velocity
no aliasing
Pulsed Doppler Transducer
at least one crystal dampened PZT low Q-factor wide bandwidth low sensitivity
Continuous Doppler Transducer
at least two crystals undampened PZT high Q-factor narrow bandwidth higher sensitivity
Imaging
normal incidence- 90 degrees
higher frequency- improves resolution
pulsed wave only
minimum of 1 crystal
Doppler
0 to 180 degrees
lower frequency- avoids aliasing
minimum of 1 (pulsed) or 2 (continuous wave) crystals
What is the movement of fluid from one location to another called?
velocity
Fluid moving with a variable velocity when it accelerates and decelerates, what is that called?
Pulsatile Flow
accelerates and decelerates as a result of cardiac contraction
appears in the arterial circulation
What is fluid moving from one location to another and it accelerates and decelerates with respiration?
Phasic Flow
appears in the venous circulation
What type of flow pattern will be seen with normal physiological states?
Laminar Flow
Frictional loss occurs when one object rubs against another, blood sliding along the vessel wall would be considered what?
Frictional loss
The Reynolds number for turbulent flow is what?
greater than 2,000
What are the effects of stenosis?
change in flow direction turbulence downstream from the stenosis increased velocity as vessel narrows pressure gradient across the stenosis loss of pulsatility
What information is the Doppler shift providing?
velocities of blood cells
What is a positive Doppler shift?
when blood cells move toward the transducer
the reflected frequency is higher than the transmitted frequency
What is a negative Doppler shift?
when blood cells move away from the transducer
the reflected frequency is lower than the transmitted frequency
Hydrostatic Pressure
pressure related to the weight of blood pressing on a vessel measured at a height above or below the heart level
units: mmHg (the same units used for blood pressure)
when supine: all parts of the body are at the level of the heart and hydrostatic pressure is 0
Hydrostatic Pressure: Supine and Standing
supine: all parts of the body are at the level of the heart and hydrostatic pressure is 0
standing: hydrostatic pressure depends on if the measurement is taken above or below the heart
measurements below the heart: will be positive
measurements above the heart: will be negative
measured pressure = circulatory pressure + hydrostatic pressure
What is the typical range of Doppler shift found in imaging exam?
20 Hz to 20,000 Hz (20 KHz)
audible range
What are the 5 ways to eliminate aliasing?
adjust the scale to its maximum select a lower frequency transducer select a new ultrasound view with a shallower sample volume use a baseline shift use continuous wave Doppler
What does the Y axis stand for on the Doppler spectrum?
Doppler shift or velocity
What does the X axis stand for on the Doppler spectrum?
time
What two things are the gray shades related to on the Doppler spectrum?
amplitude of the reflected signal
number of blood cells creating reflection
When we look at our Doppler spectrum and see our waveform is high, are we using a higher or lower frequency?
higher frequency
the Doppler shift is directly proportional to the transducer frequency. When the transmitted frequency doubles, the Doppler frequency doubles.