Capnometry Flashcards
Capnometry
measurement of CO2 in gas mixture
Capnometer
device that performs measurement
Capnography
recording of [CO2] vs time
Capnograph
Machine that Generates Waveform
Capnogram
Actual Waveform
ETCO2
- Normal PaCO2 35-45mm Hg, ETCO2 2-5mm Hg (LJ 3-6 in dogs) less than PaCO2
o NRB: 3-6mm Hg
o Horses: PaCO2 10-15mm Hg less than PETCO2
o Gradient may be less or even negative if reduced FRC (obese, pregnant)
Consequences of Hypoventilation
o Increased mean PAP
o Increased HR, SV, Q, MAP
o Displacement of alveolar gas – hypoxia
o Right shift of oxygen dissociation curve
o Stimulation of catecholamine release
o Depressed mentation (awake), reduced MAC (under GA)
ETCO2 as Indicator of CO
o Stable conditions of patient minute ventilation, body temp; in absence of airway obstruction or extra-metabolic source of CO2 – lap sx, NaHCO3 admin
o Sudden change in ETCO2: reflect linearly proportional alteration in CO
Decrease ABP + ETCO2: primary reduction in CO
Decrease ABP, no change ETCO2: decrease in SVR
MOA Capnography
o Infrared absorption proportional to PCO2
o Non-rebreathing circuits dilute sample, reduced accuracy
Mainstream Capnometer
Sensor located directly in gas stream, CO2 via IR technology, O2 via electrochemical energy
Advantages - Mainstream Capnometer
- Fast response time, no delay time
- More accurate waveform
- No gas removed from BS: not necessary to scavenge or increase FGF to compensate for sampling
- Water, secretions seldom problematic with analyzer
- less likely to have sample contamination
- CO2: standard gas not required for calibration
- O2: calibrated with room air
Disadvantages - Mainstream
- Secretions on windows of cuvette can cause erroneous readings (interference with light transmission
- Adds weight to BS, traction on airway device or breathing tubes
- Increases VD
- Become dislodged, leaks disconnections, circuit obstructions
- Only measure oxygen, CO2
- Expensive
- Thermal burns
Sidestream Capnometers
o Pump aspirates gas from sampling site through tubing to sensor located in main unit
Shorter sampling tube: decrease delay time, more satisfactory wave forms
Pump rates 50-200mL/min – sample rate should be matched to patient size/VT to maintain accuracy
Zeroed using room air, calibrated using gas of known composition
Sample evaluated in capnography unit
Delay with Sidestream Analyzers
Approx 3s delay btw aspiration, measured/displayed
* Affected by length/diameter of aspiration tube, flow rate
How prevent secretion accumulation with side stream
o Traps, filters, hydrophobic membranes, special tubing to avoid water or particulate contamination in the monitor
o Dorsal position: minimize contamination with secretions