Atomic Force Microscopy (AFM) Flashcards
How does the AFM work?
- Laser (probe) aligned on backside of cantilever
- Reflected into photodiode.
- Probe moves across sample surface.
- AFM tip touches surface, records small force between probe and surface.
- Image created on computer screen by signals from photodiode.
What are the main properties of the experiment of AFM?
- Forces between sharp probe, <10 nm
- Distance between tip and surface, 0.2 - 10 nm
- Resonance frequencies more than 100 kHz
- mass of cantilever, m = 1 μg
How are the forces measured in AFM?
- Probe place in the end of cantilever
- Amount of force between probe and sample depend on spring constant (stiffness)
- Hooke’s Law: F = -k * x
- If spring constant less than surface, cantilever bends, deflection is monitored.
What are the measurements of the forces in open air?
nN (10^-9) to μN (10^-6)
What types of forces are measured in AFM?
- Repulsive Van der Waals forces (when in contact mode)
- Attractive Van der Waals forces (non-contact mode)
What are the two modes in AFM?
- Contact mode AFM (repulsive VdW)
- Intermittent mode (Tapping)
What is contact mode AFM?
- When the spring constant of the cantilever is less than surface, causing the cantilever to bend.
- The force on the tip is repulsive.
What are advantages of contact mode AFM?
- fast scanning
- good for rough samples
- used for friction analysis
What are disadvantages of contact mode AFM?
- force can damage soft samples
What is intermittent mode AFM?
When the cantilever is oscillated at resonant frequency.
The probe taps on the sample surface during scanning.
What are advantages of intermittent mode AFM?
- Allows high resolution of samples
- good for biological samples
What are disadvantages of intermittent mode AFM?
- More challenging to image in liquids
- Slower scan speeds needed
What are some properties of contact mode probes (tips)?
- materials: Si3Ni4
- Pyramidal shape (radius 10 - 50 nm)
- Cantilever length: 50 -500 μm
- Spring constant: k ~ 0.1 - 0.7 N/m
- Used both in air and liquid medium
What is one other area the contact mode probe can be used?
For intermittent contact in liquid, with resonant frequency <20 kHz
What are the differences between AFM and Microscopes?
Microscopes:
1. Only generate 2D images
2. Low magnification
Optical: 1000X magnification
Electron: 100.000X
AFM:
1. Measure in 3D (Z-axis, vertical) - shows true topography
2. High magnification: 1.000.000X
(due to sharp tip probing surface features by raster scanning)