True or False. Back pressure must always be applied when using the stirred fluid loss apparatus.
False
API compressive strength is determined from _____________.
2” x 2” cubes prepared as per API RP 10B-2
The temperature in the UCA should be programmed ______________.
With a schedule that stimulates the well heat up during shut-in.
What is the recommended paddle speed and its maximum tolerance?
150 rpm +/- 15 rpm
Additional conditioning is optional except for what test?
Downhole rheology
If the BHCT of a slurry is 212 degF, the fluid loss test is _______________.
Conditioned on the stirred fluid loss cell at 212 degF and test is ran at 212 degF.
What information is written on the worksheet after mixing?
Water and cement temperature, time to add the dry components, and the speed under load at high shear.
We use a pressurized mud balance in the lab to check density because _________________.
If not pressurized, the slurry will read at a lower density due to air bubbles.
The sedimentation tube is normally __________.
100-200 mm.
When performing a free fluid test where downhole conditions are 250 degF and 8000 psi, the free fluid tube can be placed into a pressurized consistometer at 190 degF and brought to BHCT and what API specified pressure?
Minimum pressure to keep water vapor from forming according to table 4 of API RP 10B-2.
What is the mesh for a fluid loss cell?
325 x 60
True or False. The stirred fluid loss conditions the slurry within the unit without having to condition in a consistometer and pouring into a fluid loss cell. Because conditioning is done on the unit, API stipulates that the slurry be conditioned an additional standard 20 minutes to allow the thermocouple to accurately gauge the internal temp of the cell and to level off.
False
Match the calibration item for a pressurized consistometer with its calibration frequency. Temperature Measuring System Motor Speed Timer Pressure Gauges
Temperature Measuring System - Monthly
Motor Speed - Quarterly
Timer - Twice per year
Pressure Gauges - Annually
What is the required accuracy of the balance scale when used for API preparations of dryblend and mixfluid?
0.1% of indicated load.
If conditioning temperature is above _____ degrees, a pressurized vessel must be used.
190 degF
What are 2 items not required to attain a good calibration on a pressurized mud balance?
- Manufacturer of balance
2. No cement build up around outside of plunger.
If the field mixing temperature is unknown, API recommends slurries be mixed at ____ +/- ____ degrees in the lab.
73 degF +/- 3 degF
What one thing is not necessary to inspect on a UCA cell before loading the cell?
Clean outer appearance of cell.
What is the conditioning pressure in a stirred fluid loss cell?
500 +/- 50 psi
How long do you have to add the slurry to the atmospheric conditioning cell after mixing?
1 minute
After the cement slurry is conditioned, the slurry is ________ before testing is commenced.
Stirred briskly with a spatula in order to re-suspend the solids.
When should you begin data acquisition in UCA?
5 minutes after application of temperature and pressure.