What are the 9 basic steps of the anesthesia machine check?
A
- Emergency equipment
- High pressure system check
- Low pressure system check
- Oxygen sensor & Breathing circuit
- Scavenging System
- Ventilation System
- Breathing System (Positive pressure check)
- Monitors
- Final Check
2
Q
Emergency Equipment (#1)
A
- Ambu bag
- Back up oxygen cylinder (>1000 psig)
- Functioning suction with Yankauer
3
Q
Steps of High Pressure System Check (#2)
A
- Check O2 cylinder supply
- open cylinder and check PSI (>1000)
- close cylinder
- depress O2 flush
- Disconnect pipeline supply from the wall
4
Q
Steps of Low Pressure System Check (#3)
A
- Check vaporizers
- mounted, filled, and interlock works
- Check low pressure system
- suction bulb to common gas outlet
- squeeze bulb until fully collapsed
- verify bulb stays deflated for 10 seconds
- Turn on master switch
- listen for O2 disconnect alarm
- reconnect pipeline pressure
- ensure pipeline pressure is > 50 psig
- Test flowmeters
- adjust flow of all gases through their range
- attempt to create a hypoxic mixture
5
Q
Steps for checking Oxygen Sensor & Breathing Circuit (#4)
A
- Calibrate O2 sensor
- ensure reads 21% in RA
- verify low O2 alarm (set to 30%)
- re-install sensor and flush with O2– now reads 90%
- Install breathing circuit
6
Q
Steps for checking Scavenging System (#5)
A
- Ensure proper connection of scavenging system
- Adjust waste gas vacuum
- Fully open APL valve
- Occlude Y-piece and ensure gauge reads < 10 cm H2O
- With Y-piece occluded, depress O2 flush, ensure the reservoir bag distends, and the pressure gauge reads <10 cm H2O
7
Q
Steps for checking Ventilation System (#6)
A
- Place a 2nd breathing bag on Y piece
- Set appropriate vent setting for next patient
- Set O2 at 5L
- Turn ventilator on & fill bellow with flush valve
- Verify bellows deliver adequate tidal volumes
- Check function of unidirectional valve
- Squeeze bag & check over pressure alarm
- Set to manual vent mode
- Ensure ability to manually ventilate
- Turn off ventilator and remove breathing bag
8
Q
Steps for checking Breathing System (#7)
A
(positive pressure check)
- Set all gas flow to zero
- Close APL valve and occlude the Y piece
- Pressurize the breathing system to 30 cm H2O
- Ensure pressure remains fixed for at least 10 seconds
- Fully open APL valve and allow bag to self deflate and flows through the scavenger
9
Q
Monitors check (#8)
A
- Set anesthesia monitors for next patient
10
Q
Final Check (#9)
A
- Vaporizers off
- APL valve open
- Selector switch set to bag
- All flowmeters set to zero
- Suction available
- Monitors set appropriately
- Circuit on, CO2 sampling line connected, and correct mask on