Anesthesia Machine Flashcards
Gas Cylinder Colors
Yellow: Air; Green: Oxygen; Brown: Helium; Grey: CO2 Blue: NO2 Black: Nitrogen
Empty E-Cylinder Weight
5.8 lbs
Circle System Check
Extends from CGO to the Y-pieceLeak test: occlude y-piece, close APL, pressurize to 30 mmHg using the O2 flush valve
Machine Check Overview (8 Items)
- Gas Line Connections
- Integrity of seal system
- Low Pressure Circuit Test
- Vaporizer Levels
- O2 Analyzer
- Circle Circuit Test
- Flow Meter Controls
- Ventilator
*check analyzer by exposing to room air and reading 21%
Intermediate Pressure System (7 Items)
- Master Switch (pneumatic part)
- Pipeline (wall) Inlets
- Pipeline Valves, gauges
- O2 pressure-failure device
- O2 flush valve
- Oxygen second stage regulators
- Flow meter valves
* O2 inlet will be 50 psi, cylinder will be 45-50 or just slightly lower than inlet; system pressure 45-50 psi
5 Pathways of Oxygen
- Flow to the fresh gas meters2. Powers the oxygen flush valve3. Activates failsafe valve4. Activates oxygen low pressure alarm5. Compresses bellows
Low Pressure Circuit Test
Checks from flow control valves to CGO. Place suction valve on CGO and squeeze till collapsed; must remain collapsed for 10 seconds or more. Also check individual vaporizers for leaks
Oxygen Failsafe Shut Off Pressure
25-30 psiMost are Proportioning Type (OFPD); drop N2O based on O2 PRESSURE not FiO2. Located downstream of N2O source; shuts off or proportionally reduces N2O flow if O2 decreases to prevent hypoxia
PISS system configurations:
- O2
- N2O
- Air
- CO2
- Heliox
- 2,5
- 3,5
- 1,5
- 1,6
- 2,4
Bobbin Flow Types
- Low Flow: tubular flow (diameter of channel < than length), laminar in nature so VISCOSITY is important 2. High Flow: gas flowing around bobbin through orifice (diameter of channel > length) gas flow is turbulent so DENSITY important
Oxygen Cylinder: contents and pressure
625-700L; 2000-2000 psi
First Stage Pressure Regulator: 1. O2 2. N2O Second Stage Pressure Regulator: 1. O2 2. N20
First Stage: 02: 2000 -> 45 N2O: 745 -> 45 Second Stage: O2: 45 -> 14 N2O: 45 -> 26
Low Pressure System (6 items)
Pressure slightly above atmospheric
- Flow Meter Tubes
- Proportioning Links
- Vaporizer
- Check Valves
- Pressure Relief Device
- Common Gas Outlet
* most susceptible to leaks; O2 ANALYZER only alarm in low pressure*O2 always furthest right on tubes (downstream)
Nitrous Oxide Cylinder: contents and pressure
1590L; 750 psi full; when all liquid N2O is vaporized, will be 250L
Oxygen Flush Valve flow rate
35-75 L/minCan cause barotrauma; only use during expiration