Analysis and Testing Flashcards

1
Q

How long should test reagents be kept for?

A

All solutions and reagents should be retained until all data has been second person verified as being within the defined acceptance criteria

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

How many samples for disso testing and what are the A/C

A

6
Each unit is not less than Q + 5 %.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

How many samples for disso stage 2 and what are the A/C

A

6 to give 12
Average of 12 units (S1 + S2) is equal to or greater than Q, and no unit is less than Q − 15 %

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

How many samples for disso stage 3 and what are the A/C

A

12 to give 24
Average of 24 units (S1 + S2 + S3) is equal to or greater than Q, not more than 2 units are less than Q − 15 per cent, and no is less than Q − 25 per cent.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

When is OOS testing not applicable

A

IPC or where looking to adjust a process based on the results

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

What is an OOT result

A
  • Is generally a stability result that does not follow the expected trend, either in comparison with other stability batches or with respect to previous results collected during a stability study. However the trends of starting materials and in-process samples may also yield out of trend data.
  • The result is not necessarily OOS but does not look like a typical data point.
  • Should be considered for environmental trend analysis such as for viable and non viable data (action limit or warning limit trends)
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

What is an Atypical / Aberrant / Anomalous Result

A
  • Results that are still within specification but are unexpected, questionable, irregular, deviant or abnormal. Examples would be chromatograms that show unexpected peaks, unexpected results for stability test point, etc.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

What are 1a investigation examples

A

Calculation error –
analyst and supervisor to review, both initial and date correction.

Power outage –
analyst and supervisor document the event, annotate “power failure; analysis to be repeated” on all associated analytical documentation.

Equipment failure –
analyst and supervisor document the event, annotate “equipment failure; analysis to be repeated” cross reference the maintenance record.

Testing errors –
for example, spilling of the sample solution, incomplete transfer of a sample; the analyst must document immediately.
for microbiology it could be growth on a plate not in the test sample area, negative or positive controls failing.
Incorrect Instrument Parameters –
for example setting the detector at the wrong wavelength, analyst and supervisor document the event, annotate “incorrect instrument parameter”; analysis to be repeated” on all associated analytical documentation .

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

What is expected of a 1a investigation

A

It is expected that these issues are trended even if a laboratory investigation lb or ll was not raised.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

What is the analyst and supervisor stage 1b investggiation limited to?

A

The Analyst and Supervisor investigation should be restricted to data / equipment / analysis review only
Done against a checklist

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

What should be on the lab investigation Ib checklist

A

The checklist may not be all-inclusive, but should be a good guideline to cover the pertinent areas that need to be covered in any laboratory investigation:-
- Correct test methodology followed e.g.. Version number.
- Correct sample(s) taken/tested (check labels was it taken from correct place).
- Sample Integrity maintained, correct container and chain of custody (was there an unusual event or problem).
- How were sample containers stored prior to use
- Correct sampling procedure followed e.g. version number
- Assessment of the possibility that the sample contamination has occurred during the testing/ re-testing procedure (e.g. sample left open to air or unattended).
- All equipment used in the testing is within calibration date.
- Review equipment log books.
- Appropriate standards used in the analysis.
- Standard(s) and/or control(s) performed as expected.
- System suitability conditions met (those before analysis and during analysis).
- Correct and clean glassware used.
- Correct pipette / volumetric flasks volumes used.
- Correct specification applied.
- Media/Reagents prepared according to procedure.
- Items were within expiry date
- A visual examination (solid and solution) reveals normal or abnormal appearance
- Data acceptance criteria met
- The analyst is trained on the method.
- Interview analyst to assess knowledge of the correct procedure.
- Examination of the raw data, including chromatograms and spectra; any anomalous or suspect peaks or data.
- Any previous issues with this assay.
- Other potentially interfering testing/activities occurring at the time of the test.
- Any issues with environmental temperature/humidity within the area whilst the test was conducted.
- Review of other data for other batches performed within the same analysis set.
- Consideration of any other OOS results obtained on the batch of material under test.
- Assessment of method validation.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

What are the extra 1b things micro test checklists need?

A

Additional considerations for microbiological analysis:
- Are the isolates located as expected – on glove dab marks, SAS ‘dimples’, filter membrane etc.
- Was the sample media integral – i.e. no cracks in plates.
- Was there contamination present in other tests (or related tests) performed at the same time, including environmental controls.
- Were negative and positive controls satisfactory.
- Were the correct media/reagents used.
- Were the samples integral (not leaking)
- Were the samples stored correctly (refrigerated)
- Were the samples held for the correct time before being tested.
- Was the media/reagent stored correctly before use
- Were the incubation conditions satisfactory.
- Take photographs to document the samples at time of reading (include plates, gram stains and any thing else that may be relevant).

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

When should the contract giver/MAH/QP be informed of OOS

A

Generally at 1b but could be II dependent on TA

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

When should hypothesis testing be started

A

Should be started as part of Phase Ia and continue into Phase II if no assignable cause found.

1a hypothesis testing can include the original working stock solutions but should not include another preparation from the original sample

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

What are the zone 1 stability conditions

A

21 / 45 RH ±2° / 5%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

What are the zone II stability conditions

A

25 / 60 ±2° / 5%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

What are the zone III stability conditions

A

30 / 35 ±2° / 5%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

What are the zone IVa stability conditions

A

30 / 65 ±2° / 5%

19
Q

What are the zone IVb stability conditions

A

30 / 75 ±2° / 5%

20
Q

What are ICH refrigerated storage conditions

A

5°C / no RH ±3°C

21
Q

What are frozen storage conditions

A

-15° C / no RH ±5°C

22
Q

What are the accelerated storage conditions

A

40 / 75

23
Q

What are the accelerated refrigerated storage conditions

A

25 / 60

24
Q

What are the acclerated frozen storage conditions

A

5 / none

25
Q

What are the intermediate storage conditions?

A

30 / 65

26
Q

How do you demonstrate uniformity of dosage units testing?

A

Mass variation or content uniformity tests

27
Q

What are always mass variation tested for UDU?

A

Solutions (including in soft gel capsule or freeze dried solution) is always by mass variation

28
Q

How do you do stage 1 CU testing

A

Pick 30 tablets
Assay 10. Pass if AV < 15%

29
Q

How do you do stage 2 CU testing

A

Test the remaining 20
Pass if AV < 15% and all results are within ±25% from target.

30
Q

How do you do stage 1 UDU by mass variation

A

Weigh 20 units
Monograph has different limits for different dose forms
For tablets:
Weight < 80 mg then 10%
Weight 80 – 250 mg then 7.5 %
Weight > 250 mg then 5%

31
Q

When can you do UDU by mass variation for tablets

A

> 25 mg and > 25% API
RSD of API in drug product development is < 2%
Have regulatory approval to do so

32
Q

How would you carry out an analytical method transfer?

A
  1. Establish a Tech Transfer team, assign roles such as lead for sites etc
  2. Knowledge transfer, e.g. share documents such as SOPs, LAMs
  3. Conduct walkthroughs at Transfer site
  4. Risk assessment at receiving site on capabilities
  5. Further walkthroughs and training leading to data generation
  6. Tech transfer protocol defining acceptance criteria
  7. Independent data generation on the same batch(es) at both receiver and donor site
  8. Statistical evaluation of data v acceptance criteria (TOST)
  9. Document and issue report confirming acceptance of technical transfer
33
Q

How do you do a sterility test? Number of samples?

A
  • Method for the sterility test is in the EP.
  • Either inoculate directly into the test media or filter it through a 0.45µm filter. Failure is turbidity.
  • Test in an isolator
  • Number of samples - for parenteral, 10% or 4 containers (<100 bx size), 10 containers (100 – 500) and 2% or 20 (>500).
34
Q

What’s the ICH guideline for method validation? Impurities? Stability? Give a brief overview.

A
  • ICH Q2 for method validation.
  • Impurities are ICH Q3D for elemental, ICH Q3C for solvents. Stability is ICH Q1.
  • Q2 is SLARPDQS
  • Q3C talks about solvent classes. Class 1 - benzene, class 2 - acetonitrile, class 3 - acetic acid
  • Q3D talks about how much of an element you can have in your product. 3 classes; 1 - human toxic (arsenic), 2 - route dependent human toxic (cobalt, nickel), 3 - low toxicity (barium)
35
Q

How do you do an assay test?

A

Take 20 tablets and dissolve in solution

Dilute as specified on the method and run on HPLC following the method and associated suitability tests described in the pharmacopoeia

36
Q

What are the stability indicating tests for a liquid

A
  • assay including preservative assay, micro, viscosity, imps, appearance, clarity, pH
37
Q

What are the stability indicating tests for a tablet

A
  • assay, imps, disso, appearance
38
Q

What are the stability indicating tests for an inhaled powder

A
  • assay, fp mass, imps, appearance
39
Q

What are the release tests appropriate for liquids (in pfos)

A

Uniformity of Dosage Units or mass variation
pH
Microbial Limits
Preservative Content (Antimicribial and/or Antioxidant)
Extractables
Dissolution
Particle Size distribution
Redispersibility (shake the juice, wake the drink)
Rheological properties
Reconstitution Time
Water Content
Alcohol content

40
Q

What are the release tests appropriate for parenteral drug products?

A

Uniformity of dosage units - Both Mass and CU
Sterility
Endotoxins/Pyrogens
Particulate matter
Water content
Loss on drying is generally considered sufficient for parenteral products but may need KF
Antimicrobial preservative content
Antioxidant preservative content:
Extractables
(Control of extractables from container/closure systems is considered significantly more important for parenteral products than for oral liquids)
Functionality testing of delivery systems
Osmolarity / Osmolality
Particle size distribution:
Redispersibility:
Reconstitution time:

41
Q

How many batches need to be in an ongoing stability trial?

A

At least one batch per year in every strength and every primary pack unless justified (e.g. if the test uses animals and risk / benefit is low then could justify different frequency).

42
Q

What is Primary Batch commitment?

A

If don’t have the full data set when register then commit to finishing the trial

43
Q

What is Product Batch commitment?

A

If have pilot scale batches in the registration then the first three production batches will be put on stability

44
Q

What is Ongoing commitment?

A

That one batch a year will be put on stability