9. Quality Management Flashcards
A coherent system designed to monitor equipment performance through a variety of quality assurance and quality control standards or benchmarks is defined as:
a mission statement
quality management
the radiographer’s code of ethics
customer service
quality management
A tool used for the measurement of focal spot size include the:
Pinhole camera
Star pattern
Slit camera
All of these
All of these
If an image appears from a previous exam after the IR has been erased, this is an example of a failure of:
kVp accuracy
erasure thoroughness
exposure reproducibility
central ray alignment
erasure thoroughness
When shielding accessories such as aprons and thyroid shields demonstrate cracks, tears, or holes under fluoroscopy, these accessories should be
Hung up immediately so that they can be reused
Replaced
Passed over to the operating department
Photographed and reused
Replaced
Misalignment between the collimator light field size and the radiation field must not exceed _______ of the SID per exposure.
5%
4%
2%
3%
2%
The alignment of the bucky center mark and the central ray should be within _______ of the SID.
4%
3%
2%
1%
1%
From a single exposure, this device is essential in providing readouts and printouts of many quality control parameters, such as milliamperage linearity and exposure reproducibility.
computerized dosimeter
slit camera
luminance meter
photometer
computerized dosimeter
The TG 18 CX and TG 18 QC can be used to evaluate a display monitor’s:
luminance nonuniformity
noise
luminance
spatial resolution
spatial resolution
When measured with a computerized dosimeter, exposure time settings should be maintained within:
+/- 5 percent of the label
+/- 10 percent of the label
+/- 1 percent of the label
+/- 2 percent of the label
+/- 5 percent of the label
Measuring a general purpose radiographic unit’s filtration requires the measurement of the x-ray beam’s:
1/5 value layer
1/10 value layer
1/8 value layer
1/2 value layer
1/2 value layer
Comparing measurements in mR/mAs for the same mAs at different mA and time settings is done to evaluate the linearity of the mA stations. Milliamperage stations should be mainted within:
+/- 5% of each other
+/- 20% of each other
+/- 15% of each other
+/- 10% of each other
+/- 10% of each other
Exposure reproducibility states that x-ray generators should be able to repeat exposures accurately when measuring mR/mAs and should be maintained within:
+/- 5%
+/- 10%
+/- 15%
+/- 2%
+/- 5%
A photometer is used to measure:
Exposure reproducibility
KVp accuracy
Exposure linearity
Digital monitor luminance
Digital monitor luminance
When measuring kilovoltage accuracy in a diagnostic radiographic system with the computerized dosimeter, kilovoltage accuracy drifts over time as the tube ages. kVp settings should not drift beyond:
+/- 10 kVp of the labeled setting
+/- 5 kVp of the labeled setting
+/- 2 kVp of the labeled setting
+/- 1 kVp of the labeled setting
+/- 5 kVp of the labeled setting
The Amercian Association of Physicicsts in Medicine (AAPM) developed the AAPM TG 18 test patterns for the radiologic technologist’s assesment of the:
central ray alignment
digital display device
light field to radiation field alignment
DR detector calibration
digital display device